Image fusion of MR images and real-time ultrasonography: evaluation of fusion accuracy combining two commercial instruments, a neuronavigation system and a ultrasound system
- PMID: 15015050
- DOI: 10.1007/s00701-003-0155-6
Image fusion of MR images and real-time ultrasonography: evaluation of fusion accuracy combining two commercial instruments, a neuronavigation system and a ultrasound system
Abstract
Objective: The aim of our study was to evaluate MRI/Ultrasonography fusion accuracy depending on three ultrasonographic parameters.
Method: An ultrasonography and MRI compatible model was created, consisting of a plastic box, which contained 3 objects. MRI scans were performed with 128 sagittal slices. The objects were segmented and 3D reconstructions were created. A special ultrasound adapter with 3 reflective markers was fixed to the ultrasound probe. Thus, the probe could be tracked by the navigation system (Vector Vision(2), BrainLab, Heimstetten, Germany) and the segmented shape of the 3D-objects obtained from the MR images were overlaid onto the ultrasound display (Elegra, Siemens, Erlangen, Germany). The dependency of fusion accuracy on different depth of ultrasound display, different distances between probe and objects and different angles between the axis of the ultrasound probe and the centre of the spheres was evaluated. 435 single measurements were performed.
Findings: Overall fusion accuracy was 1.08 mm+/-0.61 mm (mean +/- standard deviation) for spheres and 1.6 mm+/-1.1 mm for arrow heads. If the ultrasound probe was directed more tangentially to the surface of the spheres the fusion became increasingly inaccurate (P<0.05). Fusion accuracy decreased the more distant the US probe was held to the object (P<0.05). Different depth of ultrasound display had no significant effect on fusion accuracy.
Conclusions: Highly accurate fusion of MR images and real-time ultrasonography could be achieved. However, careful interpretation of the fused data is necessary, when different angles and distances of the US probe to the object are concerned.
Similar articles
-
Accuracy evaluation of a 3D ultrasound-based neuronavigation system.Comput Aided Surg. 2002;7(4):197-222. doi: 10.1002/igs.10046. Comput Aided Surg. 2002. PMID: 12454892
-
Functional neuronavigation combined with intra-operative 3D ultrasound: initial experiences during surgical resections close to eloquent brain areas and future directions in automatic brain shift compensation of preoperative data.Acta Neurochir (Wien). 2007;149(4):365-78. doi: 10.1007/s00701-006-1110-0. Epub 2007 Feb 19. Acta Neurochir (Wien). 2007. PMID: 17308976
-
Real-time integration of ultrasound into neuronavigation: technical accuracy using a light-emitting-diode-based navigation system.Acta Neurochir (Wien). 2004 Nov;146(11):1211-20. doi: 10.1007/s00701-004-0352-y. Epub 2004 Sep 20. Acta Neurochir (Wien). 2004. PMID: 15375679
-
Real-time three-dimensional ultrasound methods for shape analysis and visualization.Methods. 2001 Oct;25(2):221-30. doi: 10.1006/meth.2001.1236. Methods. 2001. PMID: 11812207 Review.
-
Intra-operative 3D ultrasound in neurosurgery.Acta Neurochir (Wien). 2006 Mar;148(3):235-53; discussion 253. doi: 10.1007/s00701-005-0688-y. Epub 2005 Dec 19. Acta Neurochir (Wien). 2006. PMID: 16362178 Review.
Cited by
-
Closed-loop control in fused MR-TRUS image-guided prostate biopsy.Med Image Comput Comput Assist Interv. 2007;10(Pt 1):128-35. doi: 10.1007/978-3-540-75757-3_16. Med Image Comput Comput Assist Interv. 2007. PMID: 18051052 Free PMC article.
-
High-resolution depth electrode localization and imaging in patients with pharmacologically intractable epilepsy.J Neurosurg. 2008 Apr;108(4):812-5. doi: 10.3171/JNS/2008/108/4/0812. J Neurosurg. 2008. PMID: 18377264 Free PMC article.
-
From Grey Scale B-Mode to Elastosonography: Multimodal Ultrasound Imaging in Meningioma Surgery-Pictorial Essay and Literature Review.Biomed Res Int. 2015;2015:925729. doi: 10.1155/2015/925729. Epub 2015 May 25. Biomed Res Int. 2015. PMID: 26101779 Free PMC article. Review.
-
[Endosonographic imaging of tumor thickness in oral squamous cell cancer and its effect on the incidence of lymph node metastases].Mund Kiefer Gesichtschir. 2005 Sep;9(5):282-9. doi: 10.1007/s10006-005-0630-8. Mund Kiefer Gesichtschir. 2005. PMID: 16133281 German.
-
Value of CT-MRI fusion in iodine-125 brachytherapy for high-grade glioma.Oncotarget. 2017 Dec 1;8(68):112883-112892. doi: 10.18632/oncotarget.22844. eCollection 2017 Dec 22. Oncotarget. 2017. PMID: 29348874 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical