Improved functional vision with a modified prolate intraocular lens

J Cataract Refract Surg. 2004 May;30(5):986-92. doi: 10.1016/j.jcrs.2003.10.022.

Abstract

Purpose: To evaluate whether the Tecnis Z9000 intraocular lens (IOL) (Pfizer) with a modified prolate anterior surface provides better quality of vision than a conventional spherical IOL.

Setting: Oregon Eye Institute, Eugene, Oregon, USA.

Methods: Patients presenting for cataract surgery who were randomly assigned to receive a Tecnis Z9000 IOL (Pfizer) or a Sensar OptiEdge AR40e IOL (AMO) in 1 eye were followed for 3 months postoperatively. The patient could elect to have the same type of IOL implanted in the fellow eye. The results of sine-wave grating contrast sensitivity testing under mesopic and photopic conditions were compared interindividually.

Results: Monocular comparison was made between the 2 IOL groups, which comprised 15 patients each. The Tecnis IOL provided significantly better contrast sensitivity at 6 cycles per degree (cpd) under photopic conditions and at 1.5 and 3 cpd under mesopic conditions. Seven patients with a Tecnis IOL and 9 patients with an AR40e IOL had subsequent implantation in the fellow eye. In all eyes, including fellow eyes, having IOL implantation, the Tecnis provided significantly better contrast sensitivity at 3 and 6 cpd under photopic conditions and at 1.5, 3, and 6 cpd under mesopic conditions. The mean contrast sensitivity in fellow eyes showed that the Tecnis IOL produced significantly better results at some spatial frequencies.

Conclusions: Results show the Tecnis IOL with a modified prolate anterior surface may produce better contrast sensitivity than a standard spherical IOL under mesopic and photopic conditions. Because contrast sensitivity testing correlates well with functional vision, a goal of future research should be to evaluate patient performance using functional tests such as driving simulation.

Publication types

  • Clinical Trial
  • Comparative Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cataract Extraction
  • Contrast Sensitivity / physiology*
  • Follow-Up Studies
  • Humans
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Middle Aged
  • Prosthesis Design
  • Quality of Life
  • Vision, Ocular / physiology*