Selective determination of trimethylamine in air by liquid chromatography using solid phase extraction cartridges for sampling

J Chromatogr A. 2004 Jul 9;1042(1-2):219-23. doi: 10.1016/j.chroma.2004.05.042.

Abstract

The selective determination of trimethylamine (TMA) in air by liquid chromatography is reported. Sampling is effected by flushing air through C18-packed solid-phase extraction (SPE) cartridges at a flow rate of 15 mL/min for 15 min. Next, TMA is desorbed from the cartridges and injected into the chromatographic system. The analyte is then selectively retained on a precolumn (20 mm x 2.1 mm i.d., packed with 30 microm, Hypersil C18 phase), and derivatized on-line by injecting 9-fluorenylmethyl chloroformate (FMOC). Finally, the TMA-FMOC derivative is transferred to the analytical column (125 mm x 4 mm i.d., LiChrospher 100 RP18, 5 microm), and monitored at 262 nm. The method was applied to the measurement of TMA in air in the 0.25-2.5 microg interval (equivalent to concentrations of TMA of 1.1-11 mg/m3), providing good linearity, reproducibility and accuracy. The mean recovery of TMA was (96 +/- 7%) (n = 12), and the limit of detection was 0.05 microg. The proposed procedure allows the selective determination of TMA in the presence of other primary and secondary short-chain aliphatic amines.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air Pollutants, Occupational / analysis*
  • Boric Acids / chemistry
  • Buffers
  • Calibration
  • Chromatography, Liquid
  • Indicators and Reagents
  • Methylamines / analysis*
  • Reference Standards
  • Reproducibility of Results
  • Solutions
  • Spectrophotometry, Ultraviolet

Substances

  • Air Pollutants, Occupational
  • Boric Acids
  • Buffers
  • Indicators and Reagents
  • Methylamines
  • Solutions
  • trimethylamine
  • boric acid