Electrowetting-based microfluidics for analysis of peptides and proteins by matrix-assisted laser desorption/ionization mass spectrometry

Anal Chem. 2004 Aug 15;76(16):4833-8. doi: 10.1021/ac0498112.

Abstract

A new technique for preparing samples for matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) is reported. The technique relies on electrowetting-on-dielectric (EWOD) to move droplets containing proteins or peptides and matrix to specific locations on an array of electrodes for analysis. Standard MALDI-MS reagents, analytes, concentrations, and recipes are demonstrated to be compatible with the technique. Mass spectra are comparable to those collected by conventional methods. Nonspecific adsorption of analytes to device surfaces is demonstrated to be negligible. The results suggest that EWOD may be a useful tool for automating sample preparation for high-throughput proteomics and other applications of MALDI-MS.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Equipment Design
  • Microfluidic Analytical Techniques / instrumentation
  • Microfluidic Analytical Techniques / methods*
  • Peptides / analysis*
  • Proteins / analysis*
  • Proteome / analysis
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / instrumentation
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods

Substances

  • Peptides
  • Proteins