Simultaneous selected reaction monitoring, MS/MS and MS3 quantitation for the analysis of pharmaceutical compounds in human plasma using chip-based infusion

Rapid Commun Mass Spectrom. 2004;18(17):1995-2000. doi: 10.1002/rcm.1587.

Abstract

An assay method with mass spectrometric detection was developed for the quantitative analysis of a pharmaceutical compound and its major metabolite in human plasma using chip-based infusion. Liquid-liquid extraction sample preparation was found to be essential to minimize matrix suppression and to achieve a limit of quantitation (LOQ) of 2.5 ng/mL using a 100 microL plasma aliquot. The potential for simultaneous quantitation in selected reaction monitoring (SRM), tandem mass spectrometry (MS/MS) (enhanced product ion), and MS(3) was investigated and found to be very beneficial in improving assay selectivity. A novel concept for monitoring quantitative assay performance using a SRM/MS(3) ratio is proposed.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Mass Spectrometry / instrumentation*
  • Mass Spectrometry / methods*
  • Nanotechnology / methods*
  • Pharmaceutical Preparations / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Pharmaceutical Preparations