MR-guided vacuum-assisted breast biopsy: accuracy of targeting and success in sampling in a phantom model

Radiology. 2004 Sep;232(3):911-4. doi: 10.1148/radiol.2323031128.

Abstract

An 11-gauge MR-compatible system was designed for use in magnetic resonance vacuum-assisted breast biopsy. The system uses a detachable needle with minimal artifact to allow imaging after placement and before biopsy to confirm lesion location. A phantom study involving 16 biopsies of lesions smaller than 10 mm was conducted to assess the performance of this system. Fifteen (94%) of 16 biopsies resulted in successful lesion sampling. Eighty-five percent of all core samples contained specimen materials targeted lesion material. Further research to evaluate the efficiency, accuracy, and safety of this system in a patient population is recommended.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biopsy, Needle / instrumentation
  • Biopsy, Needle / methods*
  • Breast / pathology*
  • Equipment Design
  • Humans
  • Magnetic Resonance Imaging*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Vacuum