Subsecond chiral separations on a microchip

Electrophoresis. 2004 Nov;25(21-22):3848-52. doi: 10.1002/elps.200406028.

Abstract

Fast chiral separation of DNS-amino acids could be realized using microchip electrophoresis with fluorescence detection. For this purpose, highly sulfated cyclodextrins (HS-gamma-CD) were used as chiral selectors enabling high selectivity. Even subsecond separation of DNS-tryptophan, DNS-norleucine, DNS-phenylalanine, DNS-methionine, and DNS-aspartic acid could be achieved. Baseline separation could be accomplished within 720 ms, which is the fastest separation of enantiomers reported to date. A more complex mixture consisting of three chiral DNS-amino acids could be separated within 3.3 s utilizing a separation length of only 7 mm and an electrical field strength of 2012 V/cm.

MeSH terms

  • Amino Acids / isolation & purification
  • Cyclodextrins
  • Dansyl Compounds
  • Electrophoresis, Microchip / instrumentation
  • Electrophoresis, Microchip / methods*
  • Fluorescence
  • Optical Rotation
  • Stereoisomerism*

Substances

  • Amino Acids
  • Cyclodextrins
  • Dansyl Compounds