[Imaging characteristics of PROPELLER T2-weighted imaging]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2004 Nov;60(11):1585-91. doi: 10.6009/jjrt.kj00003326583.
[Article in Japanese]

Abstract

As the PROPELLER sequence is a combination of the radial scan and fast-spin-echo (FSE) sequence, it can be considered an FSE sequence with a motion correlation. However, there are some differences between PROPELLER and FSE owing to differences in k-space trajectory. We clarified the imaging characteristics of PROPELLER T2-weighted imaging (T2WI) for different parameters in comparison with usual FSE T2WI. When the same parameters were used, PROPELLER T2WI showed a higher signal-to-noise ratio (SNR) and lower spatial resolution than usual FSE. Effective echo time (TE) changed with different echo train lengths (ETL) or different bandwidths on PROPELLER, and imaging contrast changed accordingly to be more effective.

Publication types

  • English Abstract

MeSH terms

  • Artifacts*
  • Diffusion Magnetic Resonance Imaging / instrumentation
  • Diffusion Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging