Detrimental effects of overstuffing or understuffing with a radial head replacement in the medial collateral-ligament deficient elbow

J Bone Joint Surg Am. 2004 Dec;86(12):2629-35. doi: 10.2106/00004623-200412000-00007.

Abstract

Background: Comminuted radial head fractures associated with an injury of the medial collateral ligament can be treated with a radial head implant. We hypothesized that lengthening and shortening of the radial neck would alter the kinematics and the pressure through the radiocapitellar joint in the medial collateral ligament-deficient elbow.

Methods: The effects of lengthening (2.5 and 5 mm) and shortening (2.5 and 5 mm) of the radial neck were assessed in six human cadaveric upper extremities in which the medial collateral ligament had been surgically released. The three-dimensional spatial orientation of the ulna was recorded during simulated active motion from extension to flexion. Total varus-valgus laxity and ulnar rotation were measured. Radiocapitellar joint pressure was assessed with use of pressure-sensitive film.

Results: Radial neck lengthening or shortening of >/=2.5 mm significantly changed the kinematics in the medial collateral ligament-deficient elbow. Lengthening caused a significant decrease (p < 0.05) in varus-valgus laxity and ulnar rotation (p < 0.05), with the ulna tracking in varus and external rotation. Shortening caused a significant increase in varus-valgus laxity (p < 0.05) and ulnar rotation (p < 0.05), with the ulna tracking in valgus and internal rotation. The pressure on the radiocapitellar joint was significantly increased after 2.5 mm of lengthening.

Conclusions: This study suggests that accurate restoration of radial length is important and that axial understuffing or overstuffing of the radiohumeral joint by >/=2.5 mm alters both elbow kinematics and radiocapitellar pressure.

Clinical relevance: This in vitro cadaver study indicates that a radial head replacement should be performed with the same level of concern for accuracy and reproducibility of component position and orientation as is appropriate with any other prosthesis.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Biomechanical Phenomena
  • Cadaver
  • Collateral Ligaments / injuries*
  • Elbow Injuries*
  • Elbow Joint / physiopathology
  • Female
  • Fractures, Comminuted / surgery*
  • Humans
  • In Vitro Techniques
  • Joint Prosthesis
  • Male
  • Radius / surgery
  • Radius Fractures / surgery*
  • Range of Motion, Articular