In vitro biomechanics of cervical disc arthroplasty with the ProDisc-C total disc implant

Neurosurg Focus. 2004 Sep 15;17(3):E7.

Abstract

An in vitro biomechanical study was conducted to compare the effects of disc arthroplasty and anterior cervical fusion on cervical spine biomechanics in a multilevel human cadaveric model. Three spine conditions were studied: harvested, single-level cervical disc arthroplasty, and single-level fusion. A programmable testing apparatus was used that replicated physiological flexion/extension, lateral bending, and axial rotation. Measurements included vertebral motion, applied load, and bending moments. Relative rotations at the superior, treated, and inferior motion segment units (MSUs) were normalized with respect to the overall rotation of those three MSUs and compared using a one-way analysis of variance with Student-Newman-Keuls test (p < 0.05). Simulated fusion decreased motion across the treated site relative to the harvested and disc arthroplasty conditions. The reduced motion at the treated site was compensated at the adjacent segments by an increase in motion. For all modes of testing, use of an artificial disc prosthesis did not alter the motion patterns at either the instrumented level or adjacent segments compared with the harvested condition, except in extension.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Arthroplasty / instrumentation*
  • Arthroplasty / standards
  • Biomechanical Phenomena / instrumentation*
  • Biomechanical Phenomena / standards
  • Cervical Vertebrae* / pathology
  • Cervical Vertebrae* / surgery
  • Female
  • Humans
  • Intervertebral Disc* / pathology
  • Intervertebral Disc* / surgery
  • Male
  • Prostheses and Implants* / standards