A newly designed puncture needle for suction decompression of giant aneurysms. Technical note

J Neurosurg. 1992 May;76(5):880-2. doi: 10.3171/jns.1992.76.5.0880.

Abstract

A newly designed puncture needle for aspirating large or giant aneurysms is described. This puncture needle represents a modification of an intravenous catheter with an internal needle. It is designed to prevent blood from leaking when the internal needle is removed and has a lateral tube for aspiration. Following aneurysm puncture with the parent artery temporarily trapped, the catheter is positioned on the head frame with a brain spatula and a self-retaining retractor. Blood is suctioned through the lateral tube with a syringe or the suction system normally used in the operating room.

MeSH terms

  • Equipment Design
  • Humans
  • Intracranial Aneurysm / surgery*
  • Needles*
  • Suction / instrumentation*