Light-adjustable lens: development of in vitro nomograms

Trans Am Ophthalmol Soc. 2004:102:67-72; discussion 72-4.

Abstract

Purpose: To determine whether digital spatial intensity patterns can be developed to effect precise in vitro correction of myopic, hyperopic, and astigmatic refractive errors in a silicone light-adjustable lens (LAL). Also, to determine whether a new spatial intensity pattern for "lock-in" is effective in vitro.

Methods: A digital interferometer/irradiation system was developed to irradiate LALs and measure the power change following irradiation. Light-adjustable lenses were mounted into a wet cell maintained at 35.0 +/- 0.5 degrees C (simulated ocular temperature) and allowed to equilibrate for a minimum of 2 hours. Ultraviolet light was then applied with spatial light intensity patterns to correct hyperopia, myopia, and astigmatism. Light-adjustable lenses were also treated to effect lock-in with a separate spatial light intensity pattern. Treated lenses were characterized for power change and optical quality. In the case of lock-in, exhaustive chemical extraction was also performed to determine the percentage of remaining macromer.

Results: Appropriate digital irradiation spatial intensity patterns were created to develop nomograms for in vitro correction of myopia, hyperopia, and astigmatism in approximate 0.25 D steps. Power changes were reproducible and did not alter optical quality of the LALs. Further, lock-in dosing of the LALs did not alter optical quality or significantly change LAL power.

Conclusions: In vitro nomograms have been developed for a silicone LAL that permit precise correction of myopia, hyperopia, and astigmatism. Furthermore, a spatial light intensity pattern has been devised that effects lock-in without significantly altering LAL power or optical quality.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Astigmatism / therapy*
  • Equipment Design
  • Humans
  • Hyperopia / therapy*
  • Image Processing, Computer-Assisted
  • Interferometry / methods
  • Lens Implantation, Intraocular
  • Lenses, Intraocular*
  • Light*
  • Myopia / therapy*
  • Nomograms*
  • Optics and Photonics / instrumentation
  • Reproducibility of Results
  • Silicones
  • Ultraviolet Rays

Substances

  • Silicones