. 2005 Oct;32(10):1217-24.
doi: 10.1007/s00259-005-1823-y. Epub 2005 May 21.

A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study

Affiliations

A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study

Doron Kopelman et al. Eur J Nucl Med Mol Imaging. 2005 Oct.

Abstract

Purpose: Radioguided surgery depends on the intra-operative detection of radiolabelled tissues. This is currently accomplished with hand tools capable of providing a tone signal, depending on the proximity and direction of a radioactive source in relation to the probe. The advantages of visual images of radiolabelled tissues are well recognised, but satisfactory means of acquiring such images intra-operatively are not yet available. The goal of this study was to examine the performance of a newly developed intra-operative gamma camera, compact enough to be a hand tool and capable of yielding a visual image of the source field.

Methods: The study was performed in the laboratory with a phantom consisting of a water bath and small hollow spheres (1-2 cm in internal diameter) filled with 99mTc (1-5 microCi/cc), placed in different configurations within the bath. For comparison, studies were also performed using a standard intra-operative gamma probe, and others using a standard single-head high-resolution gamma camera.

Results: Compared with the gamma probe, the intra-operative camera was found to possess a superior ability to distinguish small, deep and weakly localised radioactivity sources from background. By acquiring images from different angles, it allowed a 3D understanding of multiple radioactive sources. It detected "cold" defects within a "hot" radiolabelled sphere. It discriminated a weak source located near a much "hotter" radioactivity source, similar to discrimination with the standard gamma camera, and discerned localised sources against a background of radioactivity.

Conclusion: It is anticipated that the high imaging potential of the camera tested in this study will offer clinical advantages.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Eur J Nucl Med. 1999 Feb;26(2):76-83 - PubMed
    1. Surg Endosc. 2001 Dec;15(12 ):1440-3 - PubMed
    1. J Nucl Med. 2001 Aug;42(8):1198-215 - PubMed
    1. Br J Surg. 2002 May;89(5):604-8 - PubMed
    1. Recent Results Cancer Res. 2000;157:273-80 - PubMed

Publication types

MeSH terms

LinkOut - more resources