A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study
- PMID: 15909193
- DOI: 10.1007/s00259-005-1823-y
A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study
Abstract
Purpose: Radioguided surgery depends on the intra-operative detection of radiolabelled tissues. This is currently accomplished with hand tools capable of providing a tone signal, depending on the proximity and direction of a radioactive source in relation to the probe. The advantages of visual images of radiolabelled tissues are well recognised, but satisfactory means of acquiring such images intra-operatively are not yet available. The goal of this study was to examine the performance of a newly developed intra-operative gamma camera, compact enough to be a hand tool and capable of yielding a visual image of the source field.
Methods: The study was performed in the laboratory with a phantom consisting of a water bath and small hollow spheres (1-2 cm in internal diameter) filled with 99mTc (1-5 microCi/cc), placed in different configurations within the bath. For comparison, studies were also performed using a standard intra-operative gamma probe, and others using a standard single-head high-resolution gamma camera.
Results: Compared with the gamma probe, the intra-operative camera was found to possess a superior ability to distinguish small, deep and weakly localised radioactivity sources from background. By acquiring images from different angles, it allowed a 3D understanding of multiple radioactive sources. It detected "cold" defects within a "hot" radiolabelled sphere. It discriminated a weak source located near a much "hotter" radioactivity source, similar to discrimination with the standard gamma camera, and discerned localised sources against a background of radioactivity.
Conclusion: It is anticipated that the high imaging potential of the camera tested in this study will offer clinical advantages.
Similar articles
-
Imaging characteristics of a 3-dimensional GSO whole-body PET camera.J Nucl Med. 2004 Jun;45(6):1040-9. J Nucl Med. 2004. PMID: 15181139
-
A hand-held imaging probe for radio-guided surgery: physical performance and preliminary clinical experience.Eur J Nucl Med Mol Imaging. 2003 Mar;30(3):339-43. doi: 10.1007/s00259-002-1064-2. Epub 2002 Dec 17. Eur J Nucl Med Mol Imaging. 2003. PMID: 12634960
-
Evaluation of fully 3-D emission mammotomography with a compact cadmium zinc telluride detector.IEEE Trans Med Imaging. 2005 Jul;24(7):868-77. doi: 10.1109/tmi.2005.852501. IEEE Trans Med Imaging. 2005. PMID: 16011316 Free PMC article.
-
Radioguided surgery: physical principles and an update on technological developments.Biomed Tech (Berl). 2020 Jan 28;65(1):1-10. doi: 10.1515/bmt-2018-0016. Biomed Tech (Berl). 2020. PMID: 31348751 Review.
-
Intraoperative Gamma Cameras: A Review of Development in the Last Decade and Future Outlook.J Imaging. 2023 May 16;9(5):102. doi: 10.3390/jimaging9050102. J Imaging. 2023. PMID: 37233321 Free PMC article. Review.
Cited by
-
Above and Beyond Robotic Surgery and 3D Modelling in Paediatric Cancer Surgery.Front Pediatr. 2021 Dec 20;9:777840. doi: 10.3389/fped.2021.777840. eCollection 2021. Front Pediatr. 2021. PMID: 34988038 Free PMC article. Review.
-
Evaluation of two intraoperative gamma detectors for assessment of 177Lu activity concentration in vivo.EJNMMI Phys. 2017 Dec;4(1):3. doi: 10.1186/s40658-016-0168-x. Epub 2017 Jan 9. EJNMMI Phys. 2017. PMID: 28070731 Free PMC article.
-
Quantitative investigation of a novel small field of view hybrid gamma camera (HGC) capability for sentinel lymph node detection.Br J Radiol. 2016 Oct;89(1066):20160435. doi: 10.1259/bjr.20160435. Epub 2016 Aug 18. Br J Radiol. 2016. PMID: 27537079 Free PMC article.
-
SPECT detectors: the Anger Camera and beyond.Phys Med Biol. 2011 Sep 7;56(17):R145-82. doi: 10.1088/0031-9155/56/17/R01. Epub 2011 Aug 9. Phys Med Biol. 2011. PMID: 21828904 Free PMC article. Review.
-
Optimization of sentinel lymph node biopsy in breast cancer using an operative gamma camera.World J Surg Oncol. 2007 Nov 17;5:132. doi: 10.1186/1477-7819-5-132. World J Surg Oncol. 2007. PMID: 18021418 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources