An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A

BJOG. 2005 Jul;112(7):915-20. doi: 10.1111/j.1471-0528.2005.00617.x.

Abstract

Objective: To assess the accuracy of an automated blood pressure device (Microlife 3BTO-A) in pregnancy and pre-eclampsia according to the British Hypertension Society (BHS) protocol.

Design: Prospective observational study.

Setting: Antenatal ward and clinics at Guy's and St Thomas' Hospital, London, UK.

Population: One hundred and five pregnant women including 35 women with non-proteinuric hypertension and 35 with pre-eclampsia.

Methods: Two trained observers took nine sequential same-arm measurements from each woman. Measurements alternated between a mercury sphygmomanometer and the device.

Main outcome measures: Grading criteria of the BHS protocol (A/B grade = pass; C/D grade = fail).

Results: The device passed the BHS protocol by achieving an A/B grade. It also achieved criteria of the Association for the Advancement of Medical Instrumentation for systolic and diastolic pressures respectively, in normotensive [-0.5 (5.7) mmHg; -0.07 (7.7) mmHg], non-proteinuric hypertensive [-3.3 (6.9) mmHg; -2.4 (6.6) mmHg] and pre-eclamptic pregnancy [-4.1 (6.4) mmHg; -1.3 (7.9) mmHg].

Conclusion: The Microlife 3BTO-A can be recommended for use in a pregnant population, including pre-eclampsia, according to the BHS protocol.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Automation
  • Blood Pressure / physiology
  • Blood Pressure Determination / instrumentation*
  • Blood Pressure Determination / standards
  • Female
  • Humans
  • Hypertension, Pregnancy-Induced / physiopathology
  • Observer Variation
  • Pre-Eclampsia / physiopathology*
  • Pregnancy
  • Sensitivity and Specificity