Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2005 Dec;31(12):1678-83; discussion 1683-4.
doi: 10.2310/6350.2005.31308.

Endovenous treatment of the great saphenous vein using a 1,320 nm Nd:YAG laser causes fewer side effects than using a 940 nm diode laser

Affiliations

Endovenous treatment of the great saphenous vein using a 1,320 nm Nd:YAG laser causes fewer side effects than using a 940 nm diode laser

Thomas M Proebstle et al. Dermatol Surg. 2005 Dec.

Abstract

Background: Limited data are available about treatment-related side effects with respect to laser wavelength in endovenous laser treatment (ELT) of the great saphenous vein (GSV).

Objective: To compare the results and side effects of a 940 nm diode and a 1,320 nm neodymium:yttium-aluminum-garnet (Nd:YAG) laser.

Methods: Three patient cohorts (A, B, and C) received ELT of the GSV using a 940 nm diode laser at 15 W (group A) or 30 W (group B) or using a 1,320 nm laser at 8 W (group C). In all cases, energy was administered continuously with constant pullback of the laser fiber under perivenous tumescent local anesthesia.

Results: The GSVs of group A (n = 113), group B (n = 136), and group C (n = 33) received ELT. An average linear endovenous energy density of 24, 63, and 62 J/cm and an average endovenous fluence equivalent of 12, 30, and 33 J/cm2 were administered to the vein. Occlusion rates were 95% (group A), 100% (group B), and 100% (group C) at day 1 after ELT and 90.3% (group A), 100% (group B), and 97% (group C) at 3 months after ELT. With the 1,320 nm laser ELT (group C), treatment-related pain (50%) and the need for analgesics (36%) were significantly reduced (p < .005) in comparison with treatment-related pain (81%) and the need for analgesics (67%) after the 30 W 940 nm laser ELT (group B). Ecchymosis was also significantly reduced (p < .05) in group C (1,320 nm) compared with group B (30 W, 940 nm).

Conclusion: ELT of the GSV using a 1,320 nm Nd:YAG laser causes fewer side effects compared with 940 nm diode laser ELT.