Optimized clinical T2 relaxometry with a standard CPMG sequence

J Magn Reson Imaging. 2006 Feb;23(2):248-52. doi: 10.1002/jmri.20490.

Abstract

Purpose: To optimize the accuracy and precision of T2 measurements using the standard Carr-Purcell-Meiboom-Gill (CPMG) sequence. T2 values obtained with this technique are normally sensitive to imperfect refocusing due to the formation of unwanted stimulated echoes.

Materials and methods: Modifications are made to the refocusing slice selection width and the interleaving scheme. A widened refocusing slice improves the uniformity of the refocusing flip angle across the slice. A slow spin echo acquisition provided "gold standard" T2 values. Repeated T2 measurements in phantom and human studies are used to compare the accuracy and precision of the optimized and non-optimized CPMG implementations.

Results: The required slice thickness ratio between refocusing and excitation slice widths is found to be 3:1 for typical optimized radiofrequency pulses. T2 values obtained using this optimized implementation more closely correspond to "gold standard" values. Repeated T2 measurements indicate significantly improved correspondence between data and model. A reduction in the fitting error of approximately 70% is demonstrated for phantoms.

Conclusion: We demonstrate that a relatively simple change to the CPMG relaxometry sequence parameters from the default setup yields significant improvements in the accuracy and precision of T2 measurements.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Artifacts
  • Electron Spin Resonance Spectroscopy
  • Female
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods
  • Male
  • Middle Aged
  • Phantoms, Imaging*
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted*
  • Software Validation