Optimized contrast-enhanced CT protocols for diagnostic whole-body 18F-FDG PET/CT: technical aspects of single-phase versus multiphase CT imaging

J Nucl Med. 2006 Mar;47(3):470-6.

Abstract

The purpose of this study was to compare various PET/CT examination protocols that use contrast-enhanced single-phase or contrast-enhanced multiphase CT scans under different breathing conditions.

Methods: Sixty patients with different malignant tumors were randomized into 4 different PET/CT protocols. Single-phase protocols included an intravenous contrast-enhanced (Ultravist 370; iodine at 370 mg/mL) single-phase whole-body CT scan (90 mL at 1.8 mL/min; delay, 90 s) during shallow breathing (protocol A) or during normal expiration (NormExp; protocol B). Multiphase protocols included 2 separate CT scans in the arterial contrast enhancement phase (90 mL at 2.5-2.8 mL/min; bolus tracking; scan range, base of the skull to the kidneys) and the portal-venous contrast enhancement phase (delay, 90 s; scan range, base of the lungs to the proximal thighs) during shallow breathing (protocol C) or during NormExp (protocol D) followed by a low-dose CT scan during shallow breathing for attenuation correction and whole-body PET. Feasibility was assessed by comparing the misalignment of the upper abdominal organs quantitatively by means of the craniocaudal, lateral, and anterior-posterior differences on coregistered PET/CT images. For image quality, the occurrence of CT artifacts and mismatching of rigid body points were evaluated qualitatively.

Results: Misalignment was significantly lower for protocol B in almost all organs and represented the best coregistration quality. Surprisingly, protocol A showed significantly better alignment than the multiphase CT scans during NormExp. Misalignment values between the multiphase protocols were not significantly different, with a trend toward lower values for protocol D. The best CT image quality, with a significantly lower occurrence of artifacts, was found for protocols B and D (NormExp). The levels of mismatching of rigid body points because of patient movement in between the transmission and emission scans were similar for all protocols.

Conclusion: Multiphase CT protocols presented a technical disadvantage represented by suboptimal image coregistration compared with single-phase protocols. Nevertheless, multiphase protocols are technically feasible and should be considered for patients who will benefit from a contrast-enhanced multiphase CT examination for diagnosis.

Publication types

  • Comparative Study
  • Randomized Controlled Trial

MeSH terms

  • Artifacts
  • Biotechnology / methods
  • Contrast Media
  • Feasibility Studies
  • Female
  • Fluorodeoxyglucose F18
  • Humans
  • Image Enhancement / methods*
  • Iohexol / analogs & derivatives*
  • Male
  • Middle Aged
  • Neoplasms / diagnosis*
  • Positron-Emission Tomography / methods*
  • Reproducibility of Results
  • Respiration*
  • Sensitivity and Specificity
  • Subtraction Technique
  • Tomography, X-Ray Computed / methods*
  • Whole Body Imaging / methods*

Substances

  • Contrast Media
  • Fluorodeoxyglucose F18
  • Iohexol
  • iopromide