Performance of the Reality polyurethane female condom and a synthetic latex prototype: a randomized crossover trial among South African women

Contraception. 2006 Apr;73(4):386-93. doi: 10.1016/j.contraception.2005.07.015. Epub 2005 Oct 19.

Abstract

Objective: This multisite, randomized, crossover trial comparing the performance of the Reality female condom (FC1) with a new synthetic latex prototype (FC2) was conducted in Durban, South Africa.

Method: In total, 276 women were enrolled and 201 women completed the study. Altogether, 1910 FC1 condoms and 1,881 FC2 condoms were used.

Results: Total breakage was 0.73% in FC1 and 0.85% in FC2 (95% confidence interval, -0.64 to 0.87). The number of clinical breakages (those that could result in a pregnancy or sexually transmitted infection) was similar for each condom type (FC1, n=9; FC2, n=8). Incorrect penetration (penis between condom and vaginal wall) was 1.26% and 0.64% for FC1 and FC2, respectively. Outer ring displacements (outer ring pushed into the vagina partially or fully) were comparable for both condoms (FC1, 3.14%; FC2, 2.98%). Slippage (condom came out of the vagina) was rare and reported in 0.37% or less of devices used. Total clinical failure was 5.24% in FC1 and 4.3% in FC2.

Conclusion: The FC1 and FC2 performed comparably within this trial.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Condoms, Female* / adverse effects
  • Cross-Over Studies
  • Equipment Failure / statistics & numerical data
  • Female
  • Humans
  • Latex*
  • Polyurethanes*
  • South Africa

Substances

  • Latex
  • Polyurethanes