Visual resolution with retinal implants estimated from recordings in cat visual cortex
- PMID: 16571357
- DOI: 10.1016/j.visres.2006.01.034
Visual resolution with retinal implants estimated from recordings in cat visual cortex
Abstract
We investigated cortical responses to electrical stimulation of the retina using epi- and sub-retinal electrodes of 20-100 microm diameter. Temporal and spatial resolutions were assessed by recordings from the visual cortex with arrays of microelectrodes and optical imaging. The estimated resolutions were approximately 40 ms and approximately 1 degrees of visual angle. This temporal resolution of 25 frames per second and spatial resolution of about 0.8 cm at about 1m and correspondingly 8 cm at 10 m distance seems sufficient for useful object recognition and visuo-motor behavior in many in- and out-door situations of daily life.
Similar articles
-
Evaluation of extraocular electrodes for a retinal prosthesis using evoked potentials in cat visual cortex.J Clin Neurosci. 2005 Jun;12(5):574-9. doi: 10.1016/j.jocn.2004.10.004. J Clin Neurosci. 2005. PMID: 16051097
-
A cortical (V1) neurophysiological recording model for assessing the efficacy of retinal visual prostheses.J Neurosci Methods. 2009 Jun 15;180(2):195-207. doi: 10.1016/j.jneumeth.2009.02.019. Epub 2009 Mar 10. J Neurosci Methods. 2009. PMID: 19464512
-
Assessing the efficacy of visual prostheses by decoding ms-LFPs: application to retinal implants.J Neural Eng. 2009 Apr;6(2):026007. doi: 10.1088/1741-2560/6/2/026007. Epub 2009 Mar 16. J Neural Eng. 2009. PMID: 19289859
-
Artificial vision: needs, functioning, and testing of a retinal electronic prosthesis.Prog Brain Res. 2009;175:317-32. doi: 10.1016/S0079-6123(09)17522-2. Prog Brain Res. 2009. PMID: 19660665 Review.
-
The retina implant--new approach to a visual prosthesis.Biomed Tech (Berl). 2004 Apr;49(4):99-103. doi: 10.1515/BMT.2004.020. Biomed Tech (Berl). 2004. PMID: 15171590 Review.
Cited by
-
Spatial Resolution of Suprachoroidal-Transretinal Stimulation Estimated by Recording Single-Unit Activity From the Cat Lateral Geniculate Nucleus.Front Neurosci. 2021 Oct 13;15:717429. doi: 10.3389/fnins.2021.717429. eCollection 2021. Front Neurosci. 2021. PMID: 34720855 Free PMC article.
-
Subretinal Transplantation of Human Embryonic Stem Cell-Derived Retinal Tissue in a Feline Large Animal Model.J Vis Exp. 2021 Aug 5;(174):10.3791/61683. doi: 10.3791/61683. J Vis Exp. 2021. PMID: 34424232 Free PMC article.
-
Response of Mouse Visual Cortical Neurons to Electric Stimulation of the Retina.Front Neurosci. 2019 Apr 4;13:324. doi: 10.3389/fnins.2019.00324. eCollection 2019. Front Neurosci. 2019. PMID: 31019449 Free PMC article.
-
Implantation of multiple suprachoroidal electrode arrays in rabbits.J Curr Ophthalmol. 2017 Dec 9;30(1):68-73. doi: 10.1016/j.joco.2017.11.007. eCollection 2018 Mar. J Curr Ophthalmol. 2017. PMID: 29564412 Free PMC article.
-
Fabrication of Pillar Shaped Electrode Arrays for Artificial Retinal Implants.Sensors (Basel). 2008 Sep 24;8(9):5845-5856. doi: 10.3390/s8095845. Sensors (Basel). 2008. PMID: 27873844 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous