Sample to sample carryover: a source of analytical laboratory error and its relevance to integrated clinical chemistry/immunoassay systems
- PMID: 16777083
- DOI: 10.1016/j.cca.2006.04.022
Sample to sample carryover: a source of analytical laboratory error and its relevance to integrated clinical chemistry/immunoassay systems
Abstract
Background: Integrated systems that combine clinical chemistry and immunoassay analyzers are used routinely. Sample to sample carryover is an inherent risk and can cause erroneously high patient test results for immunoassays. IVD manufacturers and laboratories must be aware of this phenomenon and guard against it.
Methods: We used a sample carryover protocol that directs the clinical chemistry module to process samples with very high immunoassay analyte concentrations followed by samples with very low concentrations for the same analyte. Low concentration samples were then tested by the immunoassay module to determine if the clinical chemistry module caused primary sample tube to primary sample tube carryover of the immunoassay analyte.
Results: Sample carryover was assessed on the Abbott ci8200 for HBsAg, AFP, beta-hCG, and PSA. Observed HBsAg carryover met the design specification of <0.1 ppm. Carryover for the other analytes was <0.1 ppm or below the assay limit of detection.
Conclusions: IVD manufacturers must design integrated systems to minimize primary specimen tube carryover and avoid analytical laboratory error that can impact patient safety. Carryover testing is difficult for clinical laboratories to perform in order to verify system performance. Laboratories must consider the potential for specimen carryover and its impact on results whether moving primary sample tubes between separate analyzers or using an integrated system.
Similar articles
-
Carryover: More than just a major hangover for the clinical laboratory.Clin Biochem. 2016 Jul;49(10-11):735-6. doi: 10.1016/j.clinbiochem.2016.05.020. Epub 2016 May 21. Clin Biochem. 2016. PMID: 27220058 No abstract available.
-
Evaluation of analytical methods and workflow performance of the Architect ci8200 integrated serum/plasma analyzer system.Clin Chim Acta. 2005 Jul 1;357(1):43-54. doi: 10.1016/j.cccn.2005.01.032. Epub 2005 Apr 15. Clin Chim Acta. 2005. PMID: 15963793
-
Multi-site, multi-country evaluation of analytical and operational performance of a low-mid volume chemiluminescent immunoassay analyzer.Clin Lab. 2009;55(9-10):363-70. Clin Lab. 2009. PMID: 20099573
-
Drug-testing methods and clinical interpretations of test results.Bull Narc. 1993;45(2):115-54. Bull Narc. 1993. PMID: 7920539 Review.
-
Analytical error and interference in immunoassay: minimizing risk.Ann Clin Biochem. 2011 Sep;48(Pt 5):418-32. doi: 10.1258/acb.2011.011073. Epub 2011 Jul 12. Ann Clin Biochem. 2011. PMID: 21750113 Review.
Cited by
-
Study of analytical error in lipase assay.Ann Afr Med. 2023 Jan-Mar;22(1):55-60. doi: 10.4103/aam.aam_135_21. Ann Afr Med. 2023. PMID: 36695223 Free PMC article.
-
Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination.Biochem Med (Zagreb). 2019 Jun 15;29(2):020711. doi: 10.11613/BM.2019.020711. Biochem Med (Zagreb). 2019. PMID: 31223253 Free PMC article.
-
Nutrients in Energy and One-Carbon Metabolism: Learning from Metformin Users.Nutrients. 2017 Feb 10;9(2):121. doi: 10.3390/nu9020121. Nutrients. 2017. PMID: 28208582 Free PMC article. Review.
-
Translation of proteomic biomarkers into FDA approved cancer diagnostics: issues and challenges.Clin Proteomics. 2013 Oct 2;10(1):13. doi: 10.1186/1559-0275-10-13. Clin Proteomics. 2013. PMID: 24088261 Free PMC article.
-
Steroid assays in paediatric endocrinology.J Clin Res Pediatr Endocrinol. 2010;2(1):1-16. doi: 10.4274/jcrpe.v2i1.1. Epub 2010 Feb 1. J Clin Res Pediatr Endocrinol. 2010. PMID: 21274330 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous