Iontophoresis of cyclic AMP

Biophys J. 1975 Aug;15(8):753-63. doi: 10.1016/S0006-3495(75)85852-8.

Abstract

The design, calibration, and operation of a source of controlled amounts of cyclic AMP (c-AMP) are described. Typically, 1.5 s pulses containing 10(10)-10(-12) molecules of c-AMP can be delivered to a region about 10 mum in diameter on an agar plate. The resulting concentration profiles are given as functions of distance and time. The diffusion coefficient of c-AMP in agar was measured to be 0.97 times 10(-5) cm2-s-1 at 21 degrees C.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Agar
  • Cyclic AMP*
  • Dictyostelium / metabolism
  • Diffusion
  • Electrochemistry
  • Iontophoresis / instrumentation
  • Microelectrodes

Substances

  • Agar
  • Cyclic AMP