Digital volume tomography in the diagnosis of peri-implant defects: an in vitro study on native pig mandibles

J Periodontol. 2006 Jul;77(7):1234-41. doi: 10.1902/jop.2006.050424.

Abstract

Background: The aim of this study of native pig mandibles was to investigate the accuracy and quality of the representation of peri-implant defects by intraoral radiography (IR), panoramic radiography (PR), computer tomography (CT), and digital volume tomography (DVT).

Methods: The examination was carried out on 19 native pig mandibles. In the toothless sections of the mandibles, one or two implants were inserted. Following the standardized preparation of peri-implant defects (11 each of dehiscences, fenestrations, and 2- to 3-walled intrabony defects), IR, PR, CT, and DVT were performed. The peri-implant defects were measured using appropriate software on the digitized IR and PR image programs. As a control method, the peri-implant bone defects were measured directly using a reflecting stereomicroscope with measuring ocular. The statistical comparison between the measurements of the radiographic scans and those of the direct readings of the peri-implant defects was performed with Pearson's correlation coefficient. The quality of the radiographic scans was determined through the subjective perception and detectability of the peri-implant defects by five independent observers.

Results: In the DVT and CT scans, it was possible to measure all the bone defects in three planes. Comparison with the direct peri-implant defect measurements yielded a mean deviation of 0.17+/-0.11 mm for the DVT scans and 0.18+/-0.12 mm for the CT scans. On the IR and PR images, the defects could be detected only in the mesio-distal and cranio-caudal planes. In comparison with the direct measurements of the peri-implant defects, the IR images revealed a mean deviation of 0.34+/-0.30 mm, and the PR images revealed a mean deviation of 0.41+/-0.35 mm. The quality rating of the radiographic images was highest for the DVT scans.

Conclusions: Overall, the CT and DVT scans displayed only a slight deviation in the extent of the peri-implant defects. Both radiographic imaging techniques permitted imaging of peri-implant defects in three planes, true to scale, and without overlay or distortion. The DVT scans showed the best imaging quality.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Alveolar Bone Loss / diagnostic imaging*
  • Alveolar Bone Loss / etiology
  • Animals
  • Dental Implantation, Endosseous / adverse effects
  • Dental Implants / adverse effects*
  • Imaging, Three-Dimensional
  • Mandible / diagnostic imaging
  • Radiography, Dental / methods*
  • Radiography, Dental, Digital
  • Radiography, Panoramic
  • Surgical Wound Dehiscence / diagnostic imaging
  • Swine
  • Tomography, X-Ray Computed

Substances

  • Dental Implants