Progressive increase of lithotripter output produces better in-vivo stone comminution

J Endourol. 2006 Sep;20(9):603-6. doi: 10.1089/end.2006.20.603.

Abstract

Background and purpose: Shockwave lithotripsy (SWL) has become a first-line intervention for treatment of nephrolithiasis. However, few studies have examined the effects of modifications in the method of shockwave energy administration on comminution efficiency. We propose that a gradual increase in output voltage will produce superior stone fragmentation in comparison with a constant or a decreasing output voltage by optimizing the stress wave and cavitation erosion forces on renal calculi.

Materials and methods: BegoStone phantoms were implanted in the renal pelvis of 11 pigs that underwent SWL at a pulse repetition rate of 1 Hz. Animals in the increasing strategy group (N = 4) were subjected to 18, 20, and 22 kV for 600, 600, and 800 shocks, respectively. The second group (N = 4) received a decreasing strategy of 22, 20, and 18 kV for 800, 600, and 600 shocks, respectively. The third group (N = 3) received all 2000 shocks at 20 kV, mimicking the clinical protocol.

Results: A progressively decreasing strategy and constant output voltage produced a mean comminution efficiency, or percentage of stone fragments <2 mm, of 89.0% +/- 3.3% and 87.6% +/- 1.7%, respectively. The mean comminution efficiency was improved to 96.5% +/- 1.4% by using the increasing strategy (P = 0.01).

Conclusions: A progressive increase in lithotripter output voltage during SWL can produce greater stone fragmentation than protocols employing constant or decreasing output voltage.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Disease Models, Animal
  • Female
  • Kidney / pathology
  • Kidney Calculi / pathology*
  • Kidney Calculi / therapy*
  • Lithotripsy / methods*
  • Nephrectomy
  • Random Allocation
  • Swine