Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study
- PMID: 17015786
- DOI: 10.1161/CIRCULATIONAHA.106.639450
Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study
Abstract
Background: The morbidity and mortality of surgical aortic valve replacement are increased in elderly patients with multiple high-risk comorbid conditions. Therefore, a prospective, single-center, nonrandomized study was performed in high-risk patients with aortic valve disease to evaluate the feasibility and safety of percutaneous implantation of a novel self-expanding aortic valve bioprosthesis (CoreValve).
Methods and results: Symptomatic high-risk patients with an aortic valve area <1 cm2 were considered for enrollment. CoreValve implantation was performed under general anesthesia with extracorporeal support using the retrograde approach. Clinical follow-up and transthoracic echocardiography were performed after the procedure and at days 15 and 30 after device implantation to evaluate short-term patient and device outcomes. A total of 25 patients with symptomatic aortic valve stenosis (mean gradient before implantation, 44.2+/-10.8 mm Hg) and multiple comorbidities (median logistic EuroScore, 11.0%) were enrolled. Device success and procedural success were achieved in 22 (88%) and 21 (84%) patients, respectively. Successful device implantation resulted in a marked reduction in the aortic valve gradients (mean gradient after implantation, 12.4+/-3.0 mm Hg; P<0.0001). The mean aortic regurgitation grade was unchanged. Major in-hospital cardiovascular and cerebral events occurred in 8 patients (32%), including mortality in 5 patients (20%). Among 18 patients with device success surviving to discharge, no adverse events occurred within 30 days after leaving the hospital.
Conclusions: Percutaneous implantation of the self-expanding CoreValve aortic valve prosthesis in high-risk patients with aortic stenosis with or without aortic regurgitation is feasible and, when successful, results in marked hemodynamic and clinical improvement.
Comment in
-
Letter by Grover et al regarding article, "Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg First-in-Man study".Circulation. 2007 Jun 5;115(22):e612; author reply e613. doi: 10.1161/CIRCULATIONAHA.106.675645. Circulation. 2007. PMID: 17548736 No abstract available.
Similar articles
-
Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome.J Am Coll Cardiol. 2007 Jul 3;50(1):69-76. doi: 10.1016/j.jacc.2007.04.047. Epub 2007 Jun 6. J Am Coll Cardiol. 2007. PMID: 17601548 Clinical Trial.
-
Retrograde transarterial implantation of a nonmetallic aortic valve prosthesis in high-surgical-risk patients with severe aortic stenosis: a first-in-man feasibility and safety study.Circ Cardiovasc Interv. 2008 Oct;1(2):126-33. doi: 10.1161/CIRCINTERVENTIONS.108.800607. Epub 2008 Sep 3. Circ Cardiovasc Interv. 2008. PMID: 20031667 Clinical Trial.
-
Progress and current status of percutaneous aortic valve replacement: results of three device generations of the CoreValve Revalving system.Circ Cardiovasc Interv. 2008 Dec;1(3):167-75. doi: 10.1161/CIRCINTERVENTIONS.108.819839. Circ Cardiovasc Interv. 2008. PMID: 20031675 Clinical Trial.
-
Transcatheter aortic valve implantation: anesthetic considerations.Anesth Analg. 2009 May;108(5):1453-62. doi: 10.1213/ane.0b013e31819b07ce. Anesth Analg. 2009. PMID: 19372319 Review.
-
Transcatheter aortic valve replacement in patients with critical aortic stenosis: rationale, device descriptions, early clinical experiences, and perspectives.Semin Thorac Cardiovasc Surg. 2006 Summer;18(2):165-74. doi: 10.1053/j.semtcvs.2006.08.002. Semin Thorac Cardiovasc Surg. 2006. PMID: 17157238 Review.
Cited by
-
From an Unfolding Emergency Treatment to a Universal Shift in Therapy: The History and Evolution of the Catheter-Based Treatment of Aortic Valve Stenosis.Rev Cardiovasc Med. 2022 Oct 17;23(10):349. doi: 10.31083/j.rcm2310349. eCollection 2022 Oct. Rev Cardiovasc Med. 2022. PMID: 39077143 Free PMC article. Review.
-
Four-dimensional analysis of aortic root motion in normal population using retrospective multiphase computed tomography.Eur Heart J Imaging Methods Pract. 2024 Feb 5;2(1):qyae007. doi: 10.1093/ehjimp/qyae007. eCollection 2024 Jan. Eur Heart J Imaging Methods Pract. 2024. PMID: 39045205 Free PMC article.
-
How Transcatheter Aortic Valve Implantation (TAVI) Was Born: The Struggle for a New Invention.Front Cardiovasc Med. 2021 Sep 29;8:722693. doi: 10.3389/fcvm.2021.722693. eCollection 2021. Front Cardiovasc Med. 2021. PMID: 34660724 Free PMC article. Review.
-
Application of three-dimensional transesophageal echocardiography in preoperative evaluation of transcatheter aortic valve replacement.BMC Cardiovasc Disord. 2021 Jun 28;21(1):315. doi: 10.1186/s12872-021-02101-7. BMC Cardiovasc Disord. 2021. PMID: 34182931 Free PMC article.
-
Outcomes of surgical versus transcatheter aortic valve replacement in nonagenarians- a systematic review and meta-analysis.J Community Hosp Intern Med Perspect. 2021 Jan 26;11(1):128-134. doi: 10.1080/20009666.2020.1843235. J Community Hosp Intern Med Perspect. 2021. PMID: 33552435 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources