Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study
- PMID: 17065881
- DOI: 10.1097/00000542-200611000-00008
Measurement of carboxyhemoglobin and methemoglobin by pulse oximetry: a human volunteer study
Erratum in
- Anesthesiology. 2007 Nov;107(5):863
Abstract
Background: A new eight-wavelength pulse oximeter is designed to measure methemoglobin and carboxyhemoglobin, in addition to the usual measurements of hemoglobin oxygen saturation and pulse rate. This study examines this device's ability to measure dyshemoglobins in human volunteers in whom controlled levels of methemoglobin and carboxyhemoglobin are induced.
Methods: Ten volunteers breathed 500 ppm carbon monoxide until their carboxyhemoglobin levels reached 15%, and 10 different volunteers received intravenous sodium nitrite, 300 mg, to induce methemoglobin. All were instrumented with arterial cannulas and six Masimo Rad-57 (Masimo Inc., Irvine, CA) pulse oximeter sensors. Arterial blood was analyzed by three laboratory CO-oximeters, and the resulting carboxyhemoglobin and methemoglobin measurements were compared with the corresponding pulse oximeter readings.
Results: The Rad-57 measured carboxyhemoglobin with an uncertainty of +/-2% within the range of 0-15%, and it measured methemoglobin with an uncertainty of 0.5% within the range of 0-12%.
Conclusion: The Masimo Rad-57 is the first commercially available pulse oximeter that can measure methemoglobin and carboxyhemoglobin, and it therefore represents an expansion of our oxygenation monitoring capability.
Similar articles
-
Accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.Anesth Analg. 2010 Jul;111(1):143-8. doi: 10.1213/ANE.0b013e3181c91bb6. Epub 2009 Dec 10. Anesth Analg. 2010. PMID: 20007731
-
The measurement of carboxyhemoglobin and methemoglobin using a non-invasive pulse CO-oximeter.Respir Physiol Neurobiol. 2012 Jul 1;182(2-3):88-92. doi: 10.1016/j.resp.2012.05.010. Epub 2012 May 15. Respir Physiol Neurobiol. 2012. PMID: 22609179 Clinical Trial.
-
The measurement of dyshemoglobins and total hemoglobin by pulse oximetry.Curr Opin Anaesthesiol. 2008 Dec;21(6):805-10. doi: 10.1097/ACO.0b013e328316bb6f. Curr Opin Anaesthesiol. 2008. PMID: 18997533 Review.
-
Improved accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia.Anesth Analg. 2010 Nov;111(5):1160-7. doi: 10.1213/ANE.0b013e3181f46da8. Epub 2010 Sep 14. Anesth Analg. 2010. PMID: 20841412
-
The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration.Anesth Analg. 2012 May;114(5):972-8. doi: 10.1213/ANE.0b013e318233041a. Epub 2011 Sep 29. Anesth Analg. 2012. PMID: 21965370 Review.
Cited by
-
Correlation between Carboxyhemoglobin Levels Measured by Blood Gas Analysis and by Multiwave Pulse Oximetry.J Pers Med. 2024 Jan 31;14(2):168. doi: 10.3390/jpm14020168. J Pers Med. 2024. PMID: 38392600 Free PMC article.
-
Evaluation of a Portable Blood Gas Analyzer for Prehospital Triage in Carbon Monoxide Poisoning: Instrument Validation Study.JMIR Form Res. 2023 Oct 6;7:e48057. doi: 10.2196/48057. JMIR Form Res. 2023. PMID: 37801355 Free PMC article.
-
Exploring the Anti-Hypoxaemia Effect of Hydromethylthionine: A Prospective Study of Phase 3 Clinical Trial Participants.Int J Mol Sci. 2023 Sep 6;24(18):13747. doi: 10.3390/ijms241813747. Int J Mol Sci. 2023. PMID: 37762050 Free PMC article.
-
Evaluation of methemoglobin as an intravascular contrast agent: T1 relaxation time effect in a rabbit model.Magn Reson Imaging. 2023 Nov;103:1-7. doi: 10.1016/j.mri.2023.06.018. Epub 2023 Jun 29. Magn Reson Imaging. 2023. PMID: 37392804 Free PMC article.
-
Comparison of Methemoglobin, Deoxyhemoglobin, and Ferrous Nitrosyl Hemoglobin as Potential MRI Contrast Agents.Ann Biomed Eng. 2023 Sep;51(9):2013-2020. doi: 10.1007/s10439-023-03226-w. Epub 2023 Jun 3. Ann Biomed Eng. 2023. PMID: 37270456 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources