Correlation between implant stability quotient and bone-implant contact: a retrospective histological and histomorphometrical study of seven titanium implants retrieved from humans

Clin Implant Dent Relat Res. 2006;8(4):218-22. doi: 10.1111/j.1708-8208.2006.00022.x.

Abstract

Background: Primary stability has a relevant role in the long-term success of dental implants. A quantitative method for the measurement of implant stability has been introduced (resonance frequency analysis [RFA]). Information about the significance of RFA measurements and about the relationship between RFA values and their association with implant osseointegration, success, or failure is important from a clinical point of view.

Purpose: The aim of the present histological and histomorphometric study was to see if a correlation existed between the bone-implant contact (BIC) percentage of retrieved human implants and RFA values.

Materials and methods: Seven implants inserted in the posterior mandible, with a sandblasted and acid-etched surface and retrieved after a 6-month period, were evaluated in the present study. These seven implants had been retrieved for different causes. All these implants were submerged and were retrieved with a 5-mm trephine bur and immersed in 10% buffered formalin to be processed for histology.

Results: A statistically significant correlation could be detected between implant stability quotient and BIC (p=.016).

Conclusions: Even if the relationship between bone structure and RFA is still not fully understood, in our study, a statistically significant correlation was found between RFA and BIC values. Further studies are needed to evaluate a correlation of RFA and BIC in human implants retrieved after a range of healing periods.

MeSH terms

  • Dental Implantation, Endosseous
  • Dental Implants*
  • Dental Prosthesis Retention*
  • Device Removal
  • Humans
  • Mandible
  • Osseointegration*
  • Retrospective Studies
  • Titanium
  • Vibration