High-throughput HPLC assay of acyclovir and its major impurity guanine using a monolithic column and a flow gradient approach

J Pharm Biomed Anal. 2007 Mar 12;43(4):1526-30. doi: 10.1016/j.jpba.2006.11.002. Epub 2006 Dec 4.

Abstract

Acyclovir and its major impurity guanine are determined rapidly by the incorporation of a monolithic column (100 mm x 4.6 mm i.d., Merck) to an automated HPLC system. A simple flow gradient protocol was adopted in order to accelerate the separation-detection cycle. Using 0.2% CH(3)COOH (pH 3.1) as the mobile phase and detection at 254 nm, guanine was effectively separated from the system peak (t(R)=1.25 min), while the retention time of acyclovir was 2.35 min. Linearity of the assay was validated in the range 0.1-1.0% guanine and 80-120% acyclovir (n=5). The accuracy and within- and day-to-day precision of the method were also validated, while the limits of detection and quantitation of both analytes were determined. The proposed method was successfully applied to the quality control of acyclovir raw material and the quality and stability control of acyclovir-containing pharmaceutical creams (Hagevir 5%, w/w, Cosmopharm Ltd., Korinthos, Greece).

MeSH terms

  • Acyclovir / analysis*
  • Acyclovir / chemistry
  • Antiviral Agents / analysis*
  • Antiviral Agents / chemistry
  • Automation
  • Chromatography, High Pressure Liquid / instrumentation*
  • Chromatography, High Pressure Liquid / methods*
  • Drug Contamination
  • Drug Stability
  • Guanine / analysis*
  • Guanine / chemistry
  • Molecular Structure
  • Ointments
  • Quality Control
  • Reproducibility of Results
  • Time Factors