Performance tests of two portable mini gamma cameras for medical applications
- PMID: 17153400
- DOI: 10.1118/1.2358199
Performance tests of two portable mini gamma cameras for medical applications
Abstract
We have developed two prototypes of portable gamma cameras for medical applications based on a previous prototype designed and tested by our group. These cameras use a CsI(Na) continuous scintillation crystal coupled to the new flat-panel-type multianode position-sensitive photomultiplier tube, H8500 from Hamamatsu Photonics. One of the prototypes, mainly intended for intrasurgical use, has a field of view of 44 x 44 mm2, and weighs 1.2 kg. Its intrinsic resolution is better than 1.5 mm and its energy resolution is about 13% at 140 keV. The second prototype, mainly intended for osteological, renal, mammary, and endocrine (thyroid, parathyroid, and suprarenal) scintigraphies, weighs a total of 2 kg. Its average spatial resolution is 2 mm; it has a field of view of 95 x 95 mm2, with an energy resolution of about 15% at 140 keV. The main advantages of these gamma camera prototypes with respect to those previously reported in the literature are high portability and low weight, with no significant loss of sensitivity and spatial resolution. All the electronic components are packed inside the mini gamma cameras, and no external electronic devices are required. The cameras are only connected through the universal serial bus port to a portable PC. In this paper, we present the design of the cameras and describe the procedures that have led us to choose their configuration together with the most important performance features of the cameras. For one of the prototypes, clinical tests on melanoma patients are presented and images are compared with those obtained with a conventional camera.
Similar articles
-
Design and tests of a portable mini gamma camera.Med Phys. 2004 Jun;31(6):1384-97. doi: 10.1118/1.1755570. Med Phys. 2004. PMID: 15259642
-
A micro-machined retro-reflector for improving light yield in ultra-high-resolution gamma cameras.Phys Med Biol. 2009 May 21;54(10):3003-14. doi: 10.1088/0031-9155/54/10/003. Epub 2009 Apr 23. Phys Med Biol. 2009. PMID: 19387099
-
Front-illuminated versus back-illuminated photon-counting CCD-based gamma camera: important consequences for spatial resolution and energy resolution.Phys Med Biol. 2007 Apr 21;52(8):N149-62. doi: 10.1088/0031-9155/52/8/N01. Epub 2007 Mar 23. Phys Med Biol. 2007. PMID: 17404450
-
Traditional gamma cameras are preferred.J Nucl Cardiol. 2016 Aug;23(4):795-802. doi: 10.1007/s12350-016-0460-0. Epub 2016 Apr 12. J Nucl Cardiol. 2016. PMID: 27072004 Review.
-
Nuclear probes and intraoperative gamma cameras.Semin Nucl Med. 2011 May;41(3):166-81. doi: 10.1053/j.semnuclmed.2010.12.004. Semin Nucl Med. 2011. PMID: 21440694 Review.
Cited by
-
Intraoperative Gamma Cameras: A Review of Development in the Last Decade and Future Outlook.J Imaging. 2023 May 16;9(5):102. doi: 10.3390/jimaging9050102. J Imaging. 2023. PMID: 37233321 Free PMC article. Review.
-
Above and Beyond Robotic Surgery and 3D Modelling in Paediatric Cancer Surgery.Front Pediatr. 2021 Dec 20;9:777840. doi: 10.3389/fped.2021.777840. eCollection 2021. Front Pediatr. 2021. PMID: 34988038 Free PMC article. Review.
-
Intraoperative biophotonic imaging systems for image-guided interventions.Nanophotonics. 2019 Jan;8(1):99-116. doi: 10.1515/nanoph-2018-0134. Epub 2018 Dec 14. Nanophotonics. 2019. PMID: 31187017 Free PMC article.
-
Intraoperative Imaging with a Portable Gamma Camera May Reduce the False-Negative Rate for Melanoma Sentinel Lymph Node Surgery.Ann Surg Oncol. 2018 Oct;25(11):3326-3333. doi: 10.1245/s10434-018-6685-1. Epub 2018 Aug 13. Ann Surg Oncol. 2018. PMID: 30105436 Free PMC article.
-
Development of targeted near-infrared imaging agents for prostate cancer.Mol Cancer Ther. 2014 Nov;13(11):2595-606. doi: 10.1158/1535-7163.MCT-14-0422. Epub 2014 Sep 19. Mol Cancer Ther. 2014. PMID: 25239933 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical