Experimental and clinical radiofrequency ablation: proposal for standardized description of coagulation size and geometry

Ann Surg Oncol. 2007 Apr;14(4):1381-96. doi: 10.1245/s10434-006-9033-9. Epub 2007 Jan 24.

Abstract

Background: Radiofrequency (RF) ablation is used to obtain local control of unresectable tumors in liver, kidney, prostate, and other organs. Accurate data on expected size and geometry of coagulation zones are essential for physicians to prevent collateral damage and local tumor recurrence. The aim of this study was to develop a standardized terminology to describe the size and geometry of these zones for experimental and clinical RF.

Methods: In a first step, the essential geometric parameters to accurately describe the coagulation zones and the spatial relationship between the coagulation zones and the electrodes were defined. In a second step, standard terms were assigned to each parameter.

Results: The proposed terms for single-electrode RF ablation include axial diameter, front margin, coagulation center, maximal and minimal radius, maximal and minimal transverse diameter, ellipticity index, and regularity index. In addition a subjective description of the general shape and regularity is recommended.

Conclusions: Adoption of the proposed standardized description method may help to fill in the many gaps in our current knowledge of the size and geometry of RF coagulation zones.

MeSH terms

  • Animals
  • Catheter Ablation / standards*
  • Electrodes
  • Equipment Design
  • Humans
  • Liver / pathology
  • Liver / surgery*
  • Liver Neoplasms / pathology
  • Liver Neoplasms / surgery*