Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation

Circulation. 2007 Mar 20;115(11):1392-7. doi: 10.1161/CIRCULATIONAHA.106.674259. Epub 2007 Mar 5.

Abstract

Background: We analyzed the incidence, risk factors and treatment options for stent fracture after percutaneous pulmonary valve (PPV) implantation (PPVI).

Methods and results: After PPVI, 123 patients had chest x-ray in anteroposterior and lateral projection, echocardiography, and clinical evaluation during structured follow-up. Of these 123 patients, 26 (21.1%) developed stent fracture 0 to 843 days after PPVI (stent fracture-free survival at 1 year, 85.1%; at 2 years, 74.5%; and at 3 years, 69.2%). Stent fracture was classified as type I: no loss of stent integrity (n=17); type II: loss of integrity with restenosis on echocardiography (n=8); and type III: separation of fragments or embolization (n=1). In a multivariate Cox regression, we analyzed various factors, of which 3 were associated with a higher risk of stent fracture: implantation into "native" right ventricular outflow tract (P=0.04), no calcification along the right ventricular outflow tract (judged with fluoroscopy, P=0.02), recoil of PPV (qualitatively, PPV diameter in frontal or lateral plane with fully inflated balloon > diameter after balloon deflation, P=0.03). Substernal PPV location, high-pressure post-PPVI dilatation of PPV, pre-PPVI right ventricular outflow tract gradients, and other indicators of PPV compression or asymmetry did not pose increased risk. Patients with type I fracture remain under follow-up. Patients with type II fracture had 2nd PPVI or are awaiting such procedure, and 1 patient with type III fracture required surgical explantation.

Conclusions: Stent fracture after PPVI can be managed effectively by risk stratification, systematic classification, and anticipatory management strategies. Serial x-ray and echocardiography are recommended for surveillance.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Echocardiography
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects*
  • Heart Valve Prosthesis Implantation / statistics & numerical data
  • Humans
  • Incidence
  • Male
  • Postoperative Complications / classification
  • Postoperative Complications / epidemiology
  • Postoperative Complications / therapy
  • Proportional Hazards Models
  • Prosthesis Failure*
  • Pulmonary Valve / abnormalities
  • Pulmonary Valve / surgery
  • Pulmonary Valve Insufficiency / diagnostic imaging
  • Pulmonary Valve Insufficiency / epidemiology*
  • Pulmonary Valve Insufficiency / surgery*
  • Retrospective Studies
  • Risk Factors
  • Stents / adverse effects*
  • Stents / statistics & numerical data
  • Ventricular Dysfunction, Right / diagnostic imaging
  • Ventricular Dysfunction, Right / epidemiology
  • Ventricular Dysfunction, Right / surgery