Purpose: To compare the photopic and scotopic contrast sensitivity with and without glare as well as blue color perception between eyes with an AcrySof SN60AT Natural intraocular lens (IOL) (Alcon Laboratories Inc.) and eyes with a conventional AcrySof SA60AT IOL.
Setting: Ankara University School of Medicine, Department of Ophthalmology, Ankara, Turkey.
Methods: Right eyes of 38 patients with an AcrySof Natural IOL and right eyes of 38 age-matched patients with a conventional AcrySof SA60AT IOL were included in a study. Contrast sensitivity was measured with the Functional Acuity Contrast Test under photopic conditions. Scotopic contrast sensitivity in the presence or absence of glare was measured using the Mesotest II (Oculus GmbH). Blue-green color vision was evaluated with the Moreland equation of the HMC Anomaloskop MR (Oculus GmbH).
Results: The mean age of patients was 66.6 years +/- 8.2 (SD) in the Natural IOL group and 66.4 +/- 8.0 years in the conventional IOL group. There was no statistically significant difference in photopic contrast sensitivity, scotopic contrast sensitivity with and without glare, or disability glare between the 2 groups (P>.05). Photopic and scotopic contrast sensitivity with and without glare significantly decreased with age in both groups (P<.01). There was no statistically significant difference in anomaloscope scores between the 2 groups (P>.05). The Moreland middle match point showed a significant shift toward blue with age in both groups (P<.01).
Conclusions: The AcrySof SN60AT Natural IOL provided contrast sensitivity under photopic and scotopic conditions (with and without glare) and blue color perception comparable that obtained with the AcrySof SA60AT IOL. Scotopic vision and blue color discrimination decreased with age with both IOLs.