SPE-GC/FTD determination of N-methyl-2-pyrrolidone and its metabolites in urine

J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Jul 1;854(1-2):204-10. doi: 10.1016/j.jchromb.2007.04.022. Epub 2007 Apr 25.

Abstract

An analytical method using a combination of solid-phase extraction (SPE) and gas chromatography with a flame thermionic detector (GC/FTD) was developed for determination of N-methyl-2-pyrrolidone (NMP), N-methylsuccinimide (MSI), and 2-hydroxy-N-methylsuccinimide (2-HMSI) in human urine. The SPE cartridge of poly(divinylbenzene/hydroxymethacrylate) used was directly loaded with urine sample, followed by elution with methyl isobutyl ketone (MIBK) and subsequent centrifugation, and the supernatant was injected into the capillary GC using a DB1701. This method allowed efficient separation of NMP, MSI, and 2-HMSI, which were nearly free of interference by other GC peaks arising from urine. Recoveries of NMP, MSI, and 2-HMSI from the SPE cartridge were about 98, 101, and 67%, respectively, with limits of detection of 0.04, 0.02, and 0.06 mg/L, respectively, which met the regulatory requirements. The present method was used for assay in biological monitoring of workers exposed to NMP in their occupational environment.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, Gas / methods*
  • Environmental Monitoring
  • Hydrogen-Ion Concentration
  • Pyrrolidinones / urine*
  • Sensitivity and Specificity

Substances

  • Pyrrolidinones
  • N-methylpyrrolidone