Determination of the action spectrum of the blue-light hazard for different intraocular lenses
Abstract
The spectral transmittance of various models of intraocular lenses (IOLs) has been studied. Significant differences among them were found, primarily regarding the cutoff wavelength. Based on these findings, modifications of the action spectrum for the blue-light hazard photobiological effect are proposed for each type of IOL. Moreover, the potential hazard of a representative range of radiation sources to subjects implanted with those IOLs has been calculated based on the corrected action spectra.
Similar articles
-
Violet and blue light blocking intraocular lenses: photoprotection versus photoreception.Br J Ophthalmol. 2006 Jun;90(6):784-92. doi: 10.1136/bjo.2005.086553. Br J Ophthalmol. 2006. PMID: 16714268 Free PMC article.
-
Transmission spectrums and retinal blue-light irradiance values of untinted and yellow-tinted intraocular lenses.J Cataract Refract Surg. 2010 Feb;36(2):299-307. doi: 10.1016/j.jcrs.2009.08.036. J Cataract Refract Surg. 2010. PMID: 20152614
-
The yellow intraocular lens and the natural ageing lens.Curr Opin Ophthalmol. 2014 Jan;25(1):40-3. doi: 10.1097/ICU.0000000000000020. Curr Opin Ophthalmol. 2014. PMID: 24270599 Review.
-
Ultraviolet-B phototoxicity and hypothetical photomelanomagenesis: intraocular and crystalline lens photoprotection.Am J Ophthalmol. 2010 Apr;149(4):543-9. doi: 10.1016/j.ajo.2009.11.028. Am J Ophthalmol. 2010. PMID: 20346776
-
Blue light-filtering intraocular lenses: review of potential benefits and side effects.J Cataract Refract Surg. 2009 Jul;35(7):1281-97. doi: 10.1016/j.jcrs.2009.04.017. J Cataract Refract Surg. 2009. PMID: 19545821 Review.
Cited by
-
An easy-to-apply method for determining permissible exposure limit of retina to light.Heliyon. 2022 Oct 10;8(10):e10927. doi: 10.1016/j.heliyon.2022.e10927. eCollection 2022 Oct. Heliyon. 2022. PMID: 36276735 Free PMC article.
-
Blue-blocking intraocular implants should be used routinely during phacoemulsification surgery--yes.Eye (Lond). 2012 Nov;26(11):1397-9. doi: 10.1038/eye.2012.178. Epub 2012 Sep 7. Eye (Lond). 2012. PMID: 22955039 Free PMC article. No abstract available.
-
Blue-blocking intraocular implants should be used routinely during phacoemulsification surgery--no.Eye (Lond). 2012 Nov;26(11):1400-1. doi: 10.1038/eye.2012.177. Epub 2012 Sep 7. Eye (Lond). 2012. PMID: 22955038 Free PMC article. No abstract available.
-
Measurements of transmission spectrums and estimation of retinal blue-light irradiance values of currently available clear and yellow-tinted intraocular lenses.Jpn J Ophthalmol. 2012 Jan;56(1):82-90. doi: 10.1007/s10384-011-0100-9. Epub 2011 Nov 19. Jpn J Ophthalmol. 2012. PMID: 22094397
-
Accelerated 20-year sunlight exposure simulation of a photochromic foldable intraocular lens in a rabbit model.J Cataract Refract Surg. 2011 Feb;37(2):378-85. doi: 10.1016/j.jcrs.2010.08.052. J Cataract Refract Surg. 2011. PMID: 21241924 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources