A national outbreak of Ralstonia mannitolilytica associated with use of a contaminated oxygen-delivery device among pediatric patients

Pediatrics. 2007 Jun;119(6):1061-8. doi: 10.1542/peds.2006-3739.

Abstract

Objectives: In August 2005, the Centers for Disease Control and Prevention was notified of a Ralstonia species outbreak among pediatric patients receiving supplemental oxygen therapy with the Vapotherm 2000i (Vapotherm, Inc, Stevensville, MD). The Vapotherm 2000i is a reusable medical device that was used in >900 hospitals in the United States in 2005. Ralstonia are waterborne bacilli that have been implicated in hospital-acquired infections. We initiated an investigation to determine the source of the outbreak and implement infection control and prevention measures.

Patients and methods: We performed a case-control study at 1 hospital and conducted national case findings to obtain clinical and environmental samples for laboratory analysis. Case-patients had health care-acquired Ralstonia colonization or infection. Isolates were compared by using pulsed-field gel electrophoresis. We tested manufacturer-recommended disinfection protocols for the Vapotherm 2000i under simulated-use conditions.

Results: Case-patients at the hospital (n = 5) were more likely to have received Vapotherm therapy than controls. Nationally, Ralstonia mannitolilytica was confirmed in 38 patients (aged 5 days to 7 years); 35 (92%) of the patients were exposed to the Vapotherm 2000i before recovery of the organism. Pulsed-field gel electrophoresis showed related R. mannitolilytica strains from isolates sent from 18 hospitals in 12 states. A Vapotherm machine reprocessed with a protocol proposed by the manufacturer grew Ralstonia spp after 7 days of simulated use. In December 2005, Vapotherm recalled the 2000i.

Conclusions: Our findings suggest intrinsic contamination of Vapotherm devices with Ralstonia spp. New medical devices may provide therapy equivalent to current devices yet pose novel reprocessing challenges.

Publication types

  • Comparative Study

MeSH terms

  • Case-Control Studies
  • Child
  • Child, Preschool
  • Cross Infection / epidemiology*
  • Cross Infection / etiology
  • Cross Infection / prevention & control
  • Disease Outbreaks / prevention & control
  • Equipment Contamination* / prevention & control
  • Equipment Reuse
  • Female
  • Gram-Negative Bacterial Infections / epidemiology*
  • Gram-Negative Bacterial Infections / etiology
  • Gram-Negative Bacterial Infections / prevention & control
  • Humans
  • Infant
  • Infant, Newborn
  • Male
  • Oxygen Inhalation Therapy* / adverse effects
  • Oxygen Inhalation Therapy* / instrumentation
  • Ralstonia* / isolation & purification
  • United States / epidemiology