Study objective: Using serial intravascular ultrasound (IVUS), the efficacy of reduced-dose sirolimus-eluting stents (SESs) in the prevention of neointimal hyperplasia (NH) and maintenance of luminal patency in human coronary arteries was evaluated.
Background: In the animal model, a broad therapeutic window regarding sirolimus doses in suppressing NH has been reported.
Methods: Serial cross-sectional and volumetric IVUS analyses were performed in 44 patients treated with SES that contained lower sirolimus doses (either 45% or 70%) than standard SES. For cross-sectional analysis, minimum lumen area (MLA) was measured. Percent (%) NH volumetric obstruction was calculated as 100 x NH volume/stent volume.
Results: IVUS measurements were similar between the two drug-dose groups. At 12 months follow-up, only one case developed late incomplete stent apposition. Between 4 and 12 months, a slight increase of in-stent % area loss and % NH obstruction was noted (3.5% +/- 10.4% to 6.7% +/- 10.7% and 1.9% +/- 5.0% to 4.4% +/- 8.0%, respectively). The majority of studied cases, however, sustained less than a 10% volumetric (93% of studied cases) and area loss (75% of studied cases) in the stented segment up to 12 months. At 12 months, % area loss within the stented segments and 5-mm reference segments were comparable (7.0% +/- 19.6% versus 6.7% +/- 10.7%).
Conclusions: Although slight increases of NH were noted, SESs, delivering two reduced drug doses, appeared to be effective for maintaining luminal patency during 12 months follow-up.
Copyright 2007 Wiley-Liss, Inc.