Implantable cardioverter-defibrillators and prevention of sudden cardiac death in hypertrophic cardiomyopathy
- PMID: 17652294
- DOI: 10.1001/jama.298.4.405
Implantable cardioverter-defibrillators and prevention of sudden cardiac death in hypertrophic cardiomyopathy
Erratum in
- JAMA. 2007 Oct 3;298(13):1516
Abstract
Context: Recently, the implantable cardioverter-defibrillator (ICD) has been promoted for prevention of sudden death in hypertrophic cardiomyopathy (HCM). However, the effectiveness and appropriate selection of patients for this therapy is incompletely resolved.
Objective: To study the relationship between clinical risk profile and incidence and efficacy of ICD intervention in HCM.
Design, setting, and patients: Multicenter registry study of ICDs implanted between 1986 and 2003 in 506 unrelated patients with HCM. Patients were judged to be at high risk for sudden death; had received ICDs; underwent evaluation at 42 referral and nonreferral institutions in the United States, Europe, and Australia; and had a mean follow-up of 3.7 (SD, 2.8) years. Measured risk factors for sudden death included family history of sudden death, massive left ventricular hypertrophy, nonsustained ventricular tachycardia on Holter monitoring, and unexplained prior syncope.
Main outcome measure: Appropriate ICD intervention terminating ventricular tachycardia or fibrillation.
Results: The 506 patients were predominately young (mean age, 42 [SD, 17] years) at implantation, and most (439 [87%]) had no or only mildly limiting symptoms. ICD interventions appropriately terminated ventricular tachycardia/fibrillation in 103 patients (20%). Intervention rates were 10.6% per year for secondary prevention after cardiac arrest (5-year cumulative probability, 39% [SD, 5%]), and 3.6% per year for primary prevention (5-year probability, 17% [SD, 2%]). Time to first appropriate discharge was up to 10 years, with a 27% (SD, 7%) probability 5 years or more after implantation. For primary prevention, 18 of the 51 patients with appropriate ICD interventions (35%) had undergone implantation for only a single risk factor; likelihood of appropriate discharge was similar in patients with 1, 2, or 3 or more risk markers (3.83, 2.65, and 4.82 per 100 person-years, respectively; P = .77). The single sudden death due to an arrhythmia (in the absence of advanced heart failure) resulted from ICD malfunction. ICD complications included inappropriate shocks in 136 patients (27%).
Conclusions: In a high-risk HCM cohort, ICD interventions for life-threatening ventricular tachyarrhythmias were frequent and highly effective in restoring normal rhythm. An important proportion of ICD discharges occurred in primary prevention patients who had undergone implantation for a single risk factor. Therefore, a single marker of high risk for sudden death may be sufficient to justify consideration for prophylactic defibrillator implantation in selected patients with HCM.
Republished in
-
Can ICDs prevent sudden cardiac death effectively in patients with hypertrophic cardiomyopathy?Nat Clin Pract Cardiovasc Med. 2008 Feb;5(2):76-7. doi: 10.1038/ncpcardio1062. Epub 2007 Nov 13. Nat Clin Pract Cardiovasc Med. 2008. PMID: 17998919 No abstract available.
Comment in
-
Hypertrophic cardiomyopathy, sudden death, and implantable cardiac defibrillators: how low the bar?JAMA. 2007 Jul 25;298(4):452-4. doi: 10.1001/jama.298.4.452. JAMA. 2007. PMID: 17652299 No abstract available.
-
Prognosis of apical hypertrophic cardiomyopathy.JAMA. 2007 Nov 7;298(17):2006; author reply 2006. doi: 10.1001/jama.298.17.2006-a. JAMA. 2007. PMID: 17986692 No abstract available.
Similar articles
-
Prevention of sudden cardiac death with implantable cardioverter-defibrillators in children and adolescents with hypertrophic cardiomyopathy.J Am Coll Cardiol. 2013 Apr 9;61(14):1527-35. doi: 10.1016/j.jacc.2013.01.037. J Am Coll Cardiol. 2013. PMID: 23500286
-
[The implantable cardioverter-defibrillator and hypertrophic cardiomyopathy. Experience at three centers].Rev Esp Cardiol. 2006 Jun;59(6):537-44. Rev Esp Cardiol. 2006. PMID: 16790196 Spanish.
-
Efficacy of implantable cardioverter-defibrillators for the prevention of sudden death in patients with hypertrophic cardiomyopathy.N Engl J Med. 2000 Feb 10;342(6):365-73. doi: 10.1056/NEJM200002103420601. N Engl J Med. 2000. PMID: 10666426
-
Implantable defibrillators and prevention of sudden death in hypertrophic cardiomyopathy.J Cardiovasc Electrophysiol. 2008 Oct;19(10):1118-26. doi: 10.1111/j.1540-8167.2008.01147.x. Epub 2008 Mar 31. J Cardiovasc Electrophysiol. 2008. PMID: 18384577 Review.
-
Contemporary strategies for risk stratification and prevention of sudden death with the implantable defibrillator in hypertrophic cardiomyopathy.Heart Rhythm. 2016 May;13(5):1155-1165. doi: 10.1016/j.hrthm.2015.12.048. Epub 2016 Jan 1. Heart Rhythm. 2016. PMID: 26749314 Review.
Cited by
-
Deep learning of echocardiography distinguishes between presence and absence of late gadolinium enhancement on cardiac magnetic resonance in patients with hypertrophic cardiomyopathy.Echo Res Pract. 2024 Oct 14;11(1):23. doi: 10.1186/s44156-024-00059-8. Echo Res Pract. 2024. PMID: 39396969 Free PMC article.
-
Tricuspid valve surgery following septal myectomy in patients with a cardiac implantable electronic device.JTCVS Open. 2024 Jun 27;20:29-36. doi: 10.1016/j.xjon.2024.06.011. eCollection 2024 Aug. JTCVS Open. 2024. PMID: 39296451 Free PMC article.
-
Prediction of cardiac death in patients with hypertrophic cardiomyopathy using plasma adipokine levels.Nutr Metab Cardiovasc Dis. 2024 Jun;34(6):1352-1360. doi: 10.1016/j.numecd.2024.01.017. Epub 2024 Jan 18. Nutr Metab Cardiovasc Dis. 2024. PMID: 38403486
-
Complications of Implantable Cardioverter Defibrillator and Their Potential Risk Factors in Patients with Hypertrophic Cardiomyopathy.Cardiol Res Pract. 2023 Nov 13;2023:4552100. doi: 10.1155/2023/4552100. eCollection 2023. Cardiol Res Pract. 2023. PMID: 38204930 Free PMC article.
-
Clinical course of hypertrophic cardiomyopathy patients implanted with a transvenous or subcutaneous defibrillator.Europace. 2023 Aug 2;25(9):euad270. doi: 10.1093/europace/euad270. Europace. 2023. PMID: 37724686 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical