Intraindividual comparison of high-spatial-resolution abdominal MR angiography at 1.5 T and 3.0 T: initial experience

Radiology. 2007 Sep;244(3):907-13. doi: 10.1148/radiol.2443061647.

Abstract

Purpose: To prospectively compare three-dimensional (3D) contrast material-enhanced abdominal magnetic resonance (MR) angiography at 1.5 and 3.0 T intraindividually in healthy volunteers.

Materials and methods: After institutional review board approval and informed consent were obtained, 15 healthy male volunteers (age range, 24-41 years) underwent one abdominal 3D contrast-enhanced MR angiographic examination each at 1.5 and 3.0 T in random order. Fast 3D gradient-echo sequence with parallel imaging acceleration factor of three was used for MR angiography; acquired spatial resolutions were 1x0.8x1 mm3 (imaging time, 19 seconds) at 1.5 T and 0.9x0.8x0.9 mm3 (imaging time, 18 seconds) at 3.0 T. With the latter, volume of the 3D slab was 8% larger. At 1.5 T, 20-mL bolus of gadobenate dimeglumine was delivered at 2 mL/sec; at 3.0 T, 15-mL bolus was delivered at 2.5 mL/sec. Two blinded radiologists rated image quality of aorta and proximal renal arteries in consensus with five-point scale (4=very good, 0=nondiagnostic) according to sequence and in direct intraindividual comparison. Visibility of proximal and segmental renal arteries was rated with three-point scale (3=completely visible, 1=nonvisible). Signal-to-noise ratio (SNR) was determined with phantoms. For statistical analysis of the SNRs, t tests were used.

Results: All MR angiographic measurements were diagnostic. Median score for image quality at both field strengths was 4. Depiction of proximal renal arteries was rated 3 at both field strengths. The visibility of the distal renal arteries was better at 3.0 T (median score, 3) than at 1.5 T (median score, 2). With direct comparison, 3.0-T MR angiography was better in 14 of 15 cases; no field strength was preferred in the other case. Mean SNR was significantly (P<.001) higher at 3.0 T (17.8+/-0.09 [standard deviation]) than at 1.5 T (11.9+/-0.10).

Conclusion: MR angiography at 3.0 T provided better vessel visibility and SNR than did that at 1.5 T, although voxel size and imaging time were reduced.

Publication types

  • Comparative Study

MeSH terms

  • Abdomen / blood supply*
  • Adult
  • Contrast Media / administration & dosage
  • Humans
  • Imaging, Three-Dimensional*
  • Magnetic Resonance Angiography / methods*
  • Male
  • Meglumine / administration & dosage
  • Meglumine / analogs & derivatives
  • Organometallic Compounds / administration & dosage
  • Phantoms, Imaging
  • Prospective Studies