Introduction: The purpose of this study was to analyse the applicability and advantages of the intraoperative use of a mobile isocentric C-arm with three-dimensional imaging (Siremobil ISO-C-3D) in fixation of intraarticular fractures.
Methods: After the fixation was judged to be satisfactory relying on the images provided by routine fluoroscopy, intraoperative CT visualisation with ISO-C-3D was performed to evaluate the fracture reduction and implant position. Intraoperative revision was performed based on the additional information ISO-C-3D provided beyond routine fluoroscopy. ISO-C-3D was used on a series of 72 closed-intraarticular fractures in 70 patients. Fracture distribution was: calcaneus (25), tibial plateau (17), tibial plafond (12), acetabulum (11), distal radius (3), ankle Weber-C (3) and femoral head (l). The primary outcome measure was revision rate after final ISO-C-3D data acquisition and prior to wound closure. Secondary objectives were to measure the additional time required for ISO-C-3D use and to determine the rate of further re-do surgeries.
Results: Eight out of 72 (11%) fracture fixations were judged by the surgeon to require intraoperative revision following ISO-C-3D imaging. Prior to leaving the operating room, the surgeon was satisfied with fracture alignment in all the procedures. The mean additional operative time using ISO-C-3D was 7.5 min (8.2% of the mean total operative time). No patient required re-do surgery.
Conclusion: : Intraoperative three-dimensional visualisation of intraarticular fractures enables the surgeon to identify inadvertent malreductions or implant malpositions which may be overlooked by routine C-arm fluoroscopy and hence may eliminate the need for re-do procedures. ISO-C-3D adds little operative time and may preclude the need for pre-operative and post-operative CT-scans in selected cases.