Transmission electron microscopy and atom probe specimen preparation from mechanically alloyed powder using the focused ion-beam lift-out technique

J Electron Microsc (Tokyo). 2007 Apr;56(2):43-9. doi: 10.1093/jmicro/dfm003.

Abstract

The preparation of transmission electron microscopy (TEM) and atom probe-field ion microscopy (AP-FIM) specimens from mechanically alloyed Ti-Cu-Ni-Sn powder has been explored. Applying the focused ion beam (FIB) based in situ lift-out technique, it has been demonstrated that specimen preparation can be carried on single micrometre-sized powder particles without the use of any embedding media. Since the particles did not incorporate any micropores, as revealed by cross-sectioning, the standard procedure known for bulk samples could be simply implemented to the powder material. A sequence of rectangular cuts and annular milling was found to be a highly efficient way of forming a tip-shaped AP-FIM specimen from a square cross-section blank. A Ga level < or =1 at.% was detected if a low beam current of 10 pA was chosen for the final ion-milling stages. Implanted Ga ions were mostly confined to a zone of about 2 nm in thickness and indicated that ion-induced structural transformations were negligible.

Publication types

  • Research Support, Non-U.S. Gov't