Development of a technique for the measurement of the radon exhalation rate using an activated charcoal collector

J Environ Radioact. 2008 Apr;99(4):587-95. doi: 10.1016/j.jenvrad.2007.08.024. Epub 2007 Oct 10.

Abstract

A simple system to evaluate the 222Rn (radon) exhalation rate from soil has been improved. A sampling cuvette of 2.1 L is placed so that it covers the targeted ground soil, and radon emanating from the soil accumulates within the cuvette for 24 h. Its internal radon concentration is measured by the combination of an activated charcoal (PICO-RAD) and a liquid scintillation counting system. This study shows variations of the conversion factor (CF: unit Bq m(-3)/cpm) of PICO-RAD. The range of CF due to temperature (10-30 degrees C) was between -21% and +69%, and this due to humidity (30-90%) was between 0% and -15%. Humidity and radon concentration in the cuvette covering soil tended to saturate in a few hours. The above information was used to correct the CF for the evaluation. The improved system shows high reliability and can be easily applied to natural environments.