Influence of anisotropy on peri-implant stress and strain in complete mandible model from CT
- PMID: 17951028
- DOI: 10.1016/j.compmedimag.2007.09.001
Influence of anisotropy on peri-implant stress and strain in complete mandible model from CT
Abstract
This paper reveals the influence of elastic anisotropy for the peri-implant stress and strain in personalized mandible. First, from CT data, the individual geometry of the complete range of mandible was well reproduced, also the separation between cortical and cancellous bone. Then, by an ad hoc automatic mesh generator integrated with anisotropic material assignment function, high quality anisotropic finite element model of the complete mandible was created, with two standard threaded implants embedded in posterior zone. The values of principal stress and strain in surrounding bone were evaluated under buccolingual oblique loading, and compared to that of the same FE model with equivalent isotropic material. Results of the analyses demonstrated that the percentage increase of stress and strain in anisotropic case reached up to 70%. It is concluded that anisotropy has significant effects on peri-implant stress and strain and careful consideration should be given to its use in biomechanical FE studies.
Similar articles
-
Effect of diameter and length on stress distribution of the alveolar crest around immediate loading implants.Clin Implant Dent Relat Res. 2009 Dec;11(4):279-87. doi: 10.1111/j.1708-8208.2008.00124.x. Epub 2008 Sep 9. Clin Implant Dent Relat Res. 2009. PMID: 18783411
-
Investigation of bone inelastic response in interaction phenomena with dental implants.Dent Mater. 2008 Apr;24(4):561-9. doi: 10.1016/j.dental.2007.11.024. Epub 2008 Jan 18. Dent Mater. 2008. PMID: 18207565
-
Finite element analysis of effect of prosthesis height, angle of force application, and implant offset on supporting bone.Int J Oral Maxillofac Implants. 2004 Nov-Dec;19(6):819-25. Int J Oral Maxillofac Implants. 2004. PMID: 15623056
-
Finite element stress analysis of cuneiform and cylindrical threaded implant geometries.Technol Health Care. 2006;14(4-5):421-38. Technol Health Care. 2006. PMID: 17065763 Review.
-
Biomechanics studies in dentistry: bioengineering applied in oral implantology.J Craniofac Surg. 2009 Jul;20(4):1173-7. doi: 10.1097/SCS.0b013e3181acdb81. J Craniofac Surg. 2009. PMID: 19568186 Review.
Cited by
-
A Biomechanical Analysis of the Influence of the Morfology of the Bone Blocks Grafts on the Transfer of Tension or Load to the Soft Tissue by Means of the Finite Elements Method.Materials (Basel). 2022 Dec 17;15(24):9039. doi: 10.3390/ma15249039. Materials (Basel). 2022. PMID: 36556844 Free PMC article.
-
The effect of mandibular flexure on the design of implant-supported fixed restorations of different facial types under two loading conditions by three-dimensional finite element analysis.Front Bioeng Biotechnol. 2022 Aug 29;10:928656. doi: 10.3389/fbioe.2022.928656. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 36105608 Free PMC article.
-
A high-fidelity 3D S-FEM stress analysis of a highly heterogeneous swine skull.Med Biol Eng Comput. 2020 Mar;58(3):625-641. doi: 10.1007/s11517-019-02118-3. Epub 2020 Jan 14. Med Biol Eng Comput. 2020. PMID: 31939055
-
Comparison of ultimate force revealed by compression tests on extracted first premolars and FEA with a true scale 3D multi-component tooth model based on a CBCT dataset.Clin Oral Investig. 2020 Jan;24(1):211-220. doi: 10.1007/s00784-019-02919-8. Epub 2019 May 11. Clin Oral Investig. 2020. PMID: 31079245
-
Biomechanical Effects of Various Bone-Implant Interfaces on the Stability of Orthodontic Miniscrews: A Finite Element Study.J Healthc Eng. 2017;2017:7495606. doi: 10.1155/2017/7495606. Epub 2017 Jun 19. J Healthc Eng. 2017. PMID: 29065641 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources