Diagnostic accuracy of 64-slice computed tomography coronary angiography in patients with low-to-intermediate risk

Radiol Med. 2007 Oct;112(7):969-81. doi: 10.1007/s11547-007-0198-5. Epub 2007 Oct 21.
[Article in English, Italian]

Abstract

Purpose: Our aim was to evaluate the diagnostic accuracy of 64-slice computed tomography coronary angiography (MSCT-CA) for detecting significant stenosis (>or=50% lumen reduction) in a population of patients at low to intermediate risk.

Materials and methods: We studied 72 patients (38 men, 34 women, mean age 53.9+/-8.0 years) with atypical or typical chest pain and stratified in the low-to intermediate risk category. MSCT-CA (Sensation 64 Cardiac, Siemens, Germany) was performed after IV administration of 100 ml of iodinated contrast material (Iomeprol 400 mgI/ml, Bracco, Italy). Two observers, blinded to the results of conventional coronary angiography (CAG), assessed the MSCT-CA scans in consensus. Diagnostic accuracy for detecting significant stenosis was calculated.

Results: CAG demonstrated the absence of significant disease in 70.1% of patients (51/72). No patient was excluded from MSCT-CA. There were 37 significant lesions on 1,098 available coronary segments. Sensitivity, specificity and positive and negative predictive value of MSCT-CA for detecting significant coronary artery on a per-segment basis were 100%, 98.6%, 71.2% and 100%, respectively. All patients with at least one significant lesion were correctly identified by MSCT-CA. MSCT-CA scored 15 false positives on a per-segment base, which affected only marginally the per-patient performance (only one false positive).

Conclusions: We concluded that 64-slice CT-CA is a diagnostic modality with high sensitivity and negative predictive value in patients at low to intermediate risk.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Algorithms
  • Chest Pain
  • Contrast Media
  • Coronary Angiography / methods*
  • Coronary Disease / diagnostic imaging*
  • Coronary Stenosis / diagnostic imaging
  • Data Interpretation, Statistical
  • Electrocardiography
  • Female
  • Humans
  • Iopamidol / analogs & derivatives
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Risk Factors
  • Sensitivity and Specificity
  • Tomography, Spiral Computed / methods*
  • Ventricular Function, Left

Substances

  • Contrast Media
  • iomeprol
  • Iopamidol