Diagnostic value of ultrasound in scaphoid fractures

Injury. 1991 Sep;22(5):397-9. doi: 10.1016/0020-1383(91)90104-m.

Abstract

Diagnostic ultrasound was used in 103 patients with a clinically suspected fracture of the scaphoid bone in order to achieve an early diagnosis. There were 48 women and 55 men aged 10-75 years (mean 31.4 years) tested with ultrasound. Of the patients, 72 were tested within 3 days after injury the remainder between 4 and 42 days. Fracture was confirmed radiologically in 27. We found that the ultrasound test, applied with a frequency of 1 MHz and intensity of 0.5 W/cm2 and 2.0 W/cm2 for 30s, had a sensitivity of 37 per cent and a specificity of 61 per cent. We thus conclude that ultrasound is not suitable for early diagnosis of scaphoid fracture.