Automatic determination of diltiazem and desacetyldiltiazem in human plasma using liquid-solid extraction on disposable cartridges coupled to HPLC--Part I: Optimization of the HPLC system and method validation

J Pharm Biomed Anal. 1991;9(10-12):877-82. doi: 10.1016/0731-7085(91)80017-4.

Abstract

A sensitive and automated method for the analysis of diltiazem and desacetyldiltiazem in plasma has been developed using liquid-solid extraction (LSE) on disposable extraction cartridges (DECs) in combination with HPLC. After isolation from plasma, the analytes are separated on a highly deactivated octyl silica column with a mobile phase of methanol-0.05 M phosphate buffer (pH 7.4) (62:38, v/v). The analytes are monitored photometrically at 238 nm. The complete preparation of the plasma sample as well as the injection of the final extract on to the analytical column are performed automatically by means of a sample processor equipped with a robotic arm to which is attached a needle dispensing the different liquids. The internal standard solution is first added to the plasma sample. The DEC is then conditioned successively with methanol and phosphate buffer (pH 7.4). A 1.0-ml volume of sample containing the internal standard solution is applied on an extraction cartridge filled with cyanopropyl silica (50 mg). After the DEC has been washed with the same buffer, the analytes are eluted with 0.16 ml of methanol. A 0.14-ml volume of buffer is then passed through the DEC and 0.25 ml of the final extract is injected onto the HPLC column. The absolute recoveries of the drugs are about 90% and the limit of detection for diltiazem is 0.8 ng ml-1. Relative standard deviations of 2.6% (within-day) and 3.7% (between-day) have been obtained for this compound at a plasma concentration of 50 ng ml-1.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Automation
  • Chromatography, High Pressure Liquid / instrumentation
  • Diltiazem / analogs & derivatives*
  • Diltiazem / blood*
  • Diltiazem / isolation & purification
  • Humans
  • Molecular Structure
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • deacetyldiltiazem
  • Diltiazem