Hemodynamic collapse during pulmonary embolectomy due to loss of venous return from acute occlusion of the cardiopulmonary venous cannula with thromboembolus

Interact Cardiovasc Thorac Surg. 2008 Aug;7(4):661-2. doi: 10.1510/icvts.2008.176024. Epub 2008 May 14.

Abstract

We report a case of hemodynamic collapse during pulmonary embolectomy with cardiopulmonary bypass. The hemodynamic collapse was due to complete loss of venous return due to thromboembolic occlusion of the atrial cannula. The diagnosis was made during focused intraoperative transesophageal echocardiography.

Publication types

  • Case Reports

MeSH terms

  • Adult
  • Cardiopulmonary Bypass / instrumentation*
  • Echocardiography, Transesophageal
  • Embolectomy / adverse effects*
  • Equipment Failure
  • Female
  • Hemodynamics*
  • Humans
  • Pulmonary Embolism / diagnostic imaging
  • Pulmonary Embolism / etiology
  • Pulmonary Embolism / physiopathology
  • Pulmonary Embolism / surgery*
  • Shock / diagnostic imaging
  • Shock / etiology*
  • Shock / physiopathology
  • Thromboembolism / diagnostic imaging
  • Thromboembolism / etiology*
  • Thromboembolism / physiopathology
  • Treatment Outcome
  • Ultrasonography, Interventional