Atrial fibrillation in recipients of cardiac resynchronization therapy device: 1-year results of the randomized MASCOT trial

Am Heart J. 2008 Sep;156(3):520-6. doi: 10.1016/j.ahj.2008.04.013. Epub 2008 Jul 7.

Abstract

Background: Atrial fibrillation (AF) is associated with increased morbidity and mortality in patients suffering from heart failure (HF). Patients in New York Heart Association HF classes III or IV, with systolic dysfunction and a wide QRS, are candidates for cardiac resynchronization therapy (CRT), and might benefit from atrial overdrive pacing (AOP).

Methods: The Management of Atrial fibrillation Suppression in AF-HF COmorbidity Therapy (MASCOT) trial enrolled 409 CRT device recipients (79% men), who were randomly assigned to AOP ON (n = 197), versus AOP OFF (n = 197) and followed up for 1 year. Their mean age was 68 +/- 10 years, left ventricular ejection fraction 25 +/- 6%, QRS duration 163 +/- 29 milliseconds. New York Heart Association class III was present in 86% of patients and 19% had a history of paroxysmal AF. The primary study end point was incidence of permanent AF at 1 year.

Results: Atrial overdrive pacing increased the percentage of atrial pacing from 30% to 80% (P < .0001), was well tolerated, and did not interfere with (a) delivery of CRT (95% mean ventricular pacing in both groups), (b) response to CRT (70% responders in the control vs 67% in the treatment group), or (c) cardiac function (left ventricular ejection fraction increased from 24.5% +/- 6.2% to 32.7% +/- 10.9% in the control and from 25.8% +/- 6.8% to 33.1% +/- 12.6% in the treatment group). The incidence of permanent AF was 3.3% in both groups. By logistic regression analysis, a history of AF (P < .001) and absence of antiarrhythmic drugs (P = .002) were associated with permanent AF.

Conclusions: In this first trial of a specific AF prevention algorithm in CRT recipients, AOP was safe and did not worsen HF. The prevention algorithm did not lower the 1-year incidence of AF.

Trial registration: ClinicalTrials.gov NCT00187252.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Algorithms
  • Atrial Fibrillation / epidemiology
  • Atrial Fibrillation / mortality
  • Atrial Fibrillation / prevention & control*
  • Atrial Function
  • Cardiac Pacing, Artificial / adverse effects
  • Cardiac Pacing, Artificial / methods*
  • Female
  • Follow-Up Studies
  • Heart Failure / mortality
  • Heart Failure / therapy*
  • Humans
  • Incidence
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Prostheses and Implants
  • Single-Blind Method
  • Stroke Volume
  • Treatment Outcome
  • Ventricular Function

Associated data

  • ClinicalTrials.gov/NCT00187252