Square-wave voltammetric (SWV) determination of Captopril in reconstituted serum and pharmaceutical formulations

Talanta. 2005 Feb 15;65(3):776-80. doi: 10.1016/j.talanta.2004.08.005.

Abstract

A simple, fast and sensitive square-wave voltammetric (SWV) method for the determination of trace amounts of Captopril in pharmaceutical formulation and reconstituted serum is reported. A three-electrode system containing the static mercury drop electrode (SMDE) working electrode, Pt auxiliary electrode and Ag/AgCl reference electrode was used throughout. Sodium sulfite was used as both supporting electrolyte and oxygen removing agent. No nitrogen purging is needed for oxygen removal from sample solution. Calibration graph showed good linearity in the concentration range of 0.5-50.0mugmL(-1) of Captopril and regression coefficient of 0.9957 is obtained. R.S.D. for eight replicate measurements and LOD of the proposed method are 1.2% and 6.28 x 10(-3)mugmL(-1), respectively. The effect of various parameters (equilibration time, scan increment, pulse height, drop size, frequency and sodium sulfite concentration) on the determination were investigated. The procedure was successfully applied to the determination of Captopril in pharmaceutical formulation and reconstituted serum.