Flow injection hydride generation electrothermal atomic absorption spectrometric determination of toxicologically relevant arsenic in urine

Talanta. 2006 Jul 15;69(5):1112-7. doi: 10.1016/j.talanta.2005.12.015. Epub 2006 Jan 26.

Abstract

Analytical procedure for the determination of toxicologically relevant arsenic (the sum of arsenite, arsenate, monomethylarsonate and dimethylarsinate) in urine by flow injection hydride generation and collection of generated inorganic and methylated hydrides on an integrated platform of a transverse-heated graphite atomizer for electrothermal atomic absorption spectrometric determination (ETAAS) is elaborated. Platforms are pre-treated with 2.7 micromol of zirconium and then with 0.10 micromol of iridium which serve both as an efficient hydride sequestration medium and permanent chemical modifier. Arsine, monomethylarsine and dimethylarsine are generated from diluted urine samples (10-25-fold) in the presence of 50 mmol L(-1) hydrochloric acid and 70 mmol L(-1)l-cysteine. Collection, pyrolysis and atomization temperatures are 450, 500, 2100 and 2150 degrees C, respectively. The characteristic mass, characteristic concentration and limit of detection (3sigma) are 39 pg, 0.078 microg L(-1) and 0.038 microg L(-1) As, respectively. The limits of detection in urine are ca. 0.4 and 1 microg L(-1) with 10- and 25-fold dilutions. The sample throughput rate is 25 h(-1). Applications to several urine CRMs are given.