New coil positioning method for interleaved transcranial magnetic stimulation (TMS)/functional MRI (fMRI) and its validation in a motor cortex study

J Magn Reson Imaging. 2009 Jan;29(1):189-97. doi: 10.1002/jmri.21611.

Abstract

Purpose: To develop and test a novel method for coil placement in interleaved transcranial magnetic stimulation (TMS)/functional MRI (fMRI) studies.

Materials and methods: Initially, a desired TMS coil position at the subject's head is recorded using a neuronavigation system. Subsequently, a custom-made holding device is used for coil placement inside the MR scanner. The parameters of the device corresponding to the prerecorded position are automatically determined from a fast structural image acquired directly before the experiment. The spatial accuracy of our method was verified on a phantom. Finally, in a study on five subjects, the coil was placed above the cortical representation of a hand muscle in M1 and the blood oxygenation level-dependent (BOLD) responses to short repetitive TMS (rTMS) trains were assessed using echo-planar imaging (EPI) recordings.

Results: The spatial accuracy of our method is in the range of 2.9 +/- 1.3 (SD) mm. Motor cortex stimulation resulted in robust BOLD activations in motor- and auditory related brain areas, with the activation in M1 being localized in the hand knob.

Conclusion: We present a user-friendly method for TMS coil positioning in the MR scanner that exhibits good spatial accuracy and speeds up the setup of the experiment. The motor-cortex study proves the viability of the approach and validates our interleaved TMS/fMRI setup.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Equipment Design
  • Equipment Failure Analysis
  • Evoked Potentials, Motor / physiology*
  • Humans
  • Image Enhancement / instrumentation*
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetic Resonance Imaging / methods
  • Magnetics / instrumentation*
  • Male
  • Motor Cortex / physiology*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transcranial Magnetic Stimulation / instrumentation*
  • Transcranial Magnetic Stimulation / methods