[Universal newborn hearing screening. Methodical aspects]

HNO. 2009 Jan;57(1):29-36. doi: 10.1007/s00106-008-1872-1.
[Article in German]

Abstract

Background: In order to prepare for the introduction of a universal newborn hearing screening program on a larger scale, TEOAE and ABR were recorded on automated screening instruments from both ears of 501 newborns at the University Hospital Heidelberg over a period of 13 months. The parents of children in whom OAE and ABR could not be detected in both ears, were requested to allow a complete exploration of the auditory status of the children at the department of pediatric audiology.

Subjects and methods: Internally available data networks were used for the acquisition and evaluation of data and for the organization of tracking and follow-up. Of the children 35% exhibited risk factors for congenital hearing impairment.

Results: The pass rate was 98.7% for the exclusion of binaural and 91.6% for monaural hearing disorders (TEOAE or ABR detectable). On the basis of the data it can be shown how pass rates can be optimized by selecting a suitable moment for the examination and by prescribing a minimum number of test repetitions (3 for TEOAE and 2 for AABR).

Conclusion: Quality control of screening programs should include these parameters and, in particular the number of repetitions of test measurements in all screening steps.

Publication types

  • English Abstract

MeSH terms

  • Germany
  • Hearing Disorders / diagnosis*
  • Hearing Disorders / prevention & control*
  • Hearing Tests / methods*
  • Humans
  • Infant, Newborn
  • Neonatal Screening / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity